Ground Resistance Tester Model 6471
CHAPTER 9
MAINTENANCE
9.1 Maintenance
Please make sure that you have read and fully understand the
WARNING
section on page 3.
•
To avoid electrical shock, do not attempt to perform any servicing unless
you are qualified to do so.
•
To avoid electrical shock and/or damage to the instrument, do not get water
or other foreign agents into the case.
•
Turn the instrument OFF and disconnect the unit from all circuits before
opening the case.
• Use specified spare parts only.
9.2 Cleaning
Disconnect the instrument from any source of electricity.
Use a soft cloth lightly moistened with soapy water. Wipe with a moist cloth and
then dry with a dry cloth. Never use alcohol, solvents or hydrocarbons.
9.3 Fuse Replacement
The unit is protected from overloads by two identical fuses:
• Fuse for terminal H (Z):
If this fuse is defective, the unit will no longer produce an output voltage, making it
impossible to perform resistance measurements.
To test this fuse, turn the switch to the m
Ω
position (2-Wire), connect measurement
leads to terminals
H (Z)
and
E (X)
and start a measurement. If the unit refuses to
operate and the symbol for terminal
H (Z)
blinks, the fuse must be replaced.