Ground Resistance Tester Model 6471
Press the ► button to choose the following frequencies:
•
USr, 55, 92, 110, 119, 128 Hz (128 Hz is default).
To change the user frequency (USr), press the ▲▼ buttons (▲▼ raises the value,
and
+ ▲▼ lowers it).
Table of possible USr frequencies (91 values from 41 to 513Hz):
41
43
46
49
50
55
60
61
64
67
69
73
79
82
85
92
98
101
110
119
122
128
134
137
146
159
165
171
183
195
201
220
238
244
256
269
275
293
317
330
342
366
391
403
439
476
488
513
There are two user frequencies: one for resistivity and one for earth and soil po
-
tential measurements. These two values remain in memory even after the device
is switched off.
For soil resistivity measurements, the user frequency is limited to 128 Hz.
6.3 Manual Settings for m
Ω
Measurements
Pressing
Hz/OPTIONS
in
MANUAL
mode allows the following parameters to be
changed using the ► button:
•
Terminal symbols H and E blink
H S ES E blink (4-Pole measurement)
•
POS on H and DC+ blinks
neg on H and DC- (reversal of polarity at terminal H)
In
MANUAL
mode the tester does not reverse polarity automatically during the
measurement; however, the polarity can be reversed during the measurement by
pressing the
Hz/OPTIONS
button.
6.3.1 Continuity Test
The 2-wire m
Ω
measurement gives a quick measurement result, accompanied by
an audible beep, during a continuity check.
The display is in the sensitivity range (0.5
Ω
to 1.99k
Ω
) and the terminal check is
limited to terminal
H (Z)
(a cable must be connected to it), making it possible to
start the measurement with an open circuit.
For a continuity check, the following settings are mandatory:
• The 2-wire m
Ω
measurement function must be selected.
•
The device must be in manual mode.
•
The alarm function must be active (On).
•
The alarm threshold must be low (<).
•
The buzzer must be activated (bEEP On).