Ground Resistance Tester Model 6471
45
CHAPTER 6
MANUAL MODE
All of the measurement functions described in § 5 (
AUTO mode
) can be performed
in
MANUAL
mode if necessary.
To access
MANUAL
mode:
•
Press the
Hz/OPTIONS
button once. The
Hz/OPTIONS
symbol appears
and the
AUTO
indicator blinks.
• Use the ► button to switch from
AUTO
to
MANUAL
.
• When the tester is in
MANUAL
mode you can press
Hz/OPTIONS
repeatedly
to see various parameters that can be set, depending on the measurement
function selected.
When a measurement is performed by a short or long press on the
START/STOP
button (the circular arrows
on the display will rotate), the measurement can be
stopped by a second press of this button.
Whenever a new measurement function is selected, the device automatically
changes back to
AUTO
mode.
6.1 Switching the Test Voltage
Available in all functions except m
Ω
and 2 Clamp.
If necessary, the test voltages can be switched to either 16 or 32V as follows:
•
Press the
Hz/OPTIONS
button, then press it again to make the output volt
-
age (Uout) blink.
• Use the ► button to switch to 16 or 32V, then press
Hz/OPTIONS
again.
This setting is retained when the unit is turned off.
6.2 Measurement Frequency Selection
If there are signals with a frequency that could interfere with the chosen test fre
-
quency, the
NOISE
symbol will appear.
While in
MANUAL
mode, change the frequency by pressing the
Hz/OPTIONS
button until the frequency flashes.
www.
ShopAEMC
.com
Shop for AEMC products online at:
1.888.610.7664