14-8
Art: 714376-00M
Rev. Date: 31-Jan-12
Ranges
The ranges displayed represent the maximum deviation expected when controls
and cartridges are performing properly.
Should results outside the ranges be obtained, refer to the Troubleshooting
section that follows the Procedure for Testing Controls.
Target Values are specific to the i-STAT System. Results obtained from these
aqueous controls with other methods may differ due to sample matrix effects.
RNA
®
MEDICAL HEMATOCRIT CONTROL
Intended Use
RNA Medical Hematocrit Control is a quality control material used for verifying
the performance of hematocrit sensors in i-STAT cartridges.
Product Description
Hematocrit controls are provided in three (3) levels for monitoring cartridge
performance at different points within the clinical range. The controls are
packaged in sealed glass ampules, each containing 1.7 mL of solution. Ampules
are packaged ten (10) per box.
Active Ingredients
This product is a buffered aqueous solution containing electrolytes and non-
conductive ingredients. This control contains no red blood cells, and no human
or biological material.
Storage
RNA Medical Hematocrit Control is stable until the expiration date stated on
the ampule when stored at temperatures of 2-25 °C. Do not freeze or expose
ampules to temperatures greater than 30 ºC.
If stored refrigerated, the control material should be equilibrated to room
temperature for at least four (4) hours prior to testing.
The product may turn slightly yellow in color over its shelf life if stored at
room temperature. This coloration is normal and does not affect product
performance.
Directions For Use
For each lot of i-STAT cartridges received, use a representational number of
cartridges to analyze multiple levels of RNA Medical Hematocrit controls.
1. Gently invert the ampule to mix the solution. Tap the ampule to
restore the liquid to the bottom of the ampule.
2. Open the ampule by snapping off the tip at the neck. Use gauze,
tissue, gloves, or an appropriate ampule opener to protect fingers
from cuts.
3. Fill and seal a cartridge and insert immediately into the analyzer.
i. Note: The Control option from the Quality Tests menu must
be used on the i-STAT 1 Analyzer for RNA Medical Hematocrit
Control.
4. Compare the i-STAT System Hematocrit result to the value
assignment sheet value. If the result is within the expected range,
use the cartridges as needed.
i. If the results are outside of the expected range, consult page 14-19
of this section for troubleshooting recommendations.
Содержание i-STAT 1
Страница 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Страница 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Страница 44: ......
Страница 90: ......
Страница 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Страница 162: ......
Страница 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Страница 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Страница 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Страница 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Страница 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Страница 192: ......
Страница 194: ......
Страница 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 226: ......
Страница 236: ......
Страница 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Страница 254: ......
Страница 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Страница 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Страница 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Страница 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Страница 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Страница 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Страница 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Страница 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Страница 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Страница 536: ......
Страница 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Страница 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Страница 572: ...Art 715617 00D Rev Date 06 11 08 ...
Страница 578: ...Art 714547 00F Rev Date 06 11 08 ...
Страница 590: ...Art 720735 00B Rev Date 06 11 08 ...
Страница 592: ...Art 721106 00B Rev Date 06 11 08 ...
Страница 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Страница 610: ...Art 720654 00B Rev Date 06 11 08 ...
Страница 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Страница 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Страница 618: ...Art 721296 00A Rev Date 07 16 08 ...
Страница 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Страница 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 658: ......