14-10
Art: 714376-00M
Rev. Date: 31-Jan-12
Control Target Values
and Expected Ranges
Target values (determined by testing multiple vials of each level using
multiple lots of i-STAT cartridges with analyzers that have passed the
Electronic Simulator test) are printed on a value assignment sheet
posted on the APOC website at www.abbottpointofcare.com. The ranges
displayed represent the maximum deviation expected when controls
and cartridges are performing properly. Should results outside the range
be obtained, refer to the Troubleshooting portion of this section of the
i-STAT System Manual. Always be sure that the lot number printed on the
value assignment sheet matches the lot number on the label of the vial in use,
and that the software revision above the table matches the software revision
in the analyzer (check the status page on the analyzer).
Note
: Target values are specific to the i-STAT System; results obtained from these
reconstituted control plasmas may differ if used with other methods.
STEP
ACTION
1
After 45 minute room temperature equilibration, remove the cap
and stopper from one lyophilized human plasma control vial and
remove the cap from one vial of calcium chloride reconstituting
fluid.
2
Pour the entire contents of the calcium chloride vial into the
lyophilized human plasma control vial. Place the stopper back
in the reconstituted control vial, sealing the vial appropriately so
that the contents do not leak or spill out.
3
Allow the vial to sit at room temperature for 1 minute.
4
Mix the contents of the vial by swirling gently for 1
m i n u t e , t h e n i n v e r t i n g s l o w l y f o r 3 0 s e c o n d s .
Note:
To minimize foaming of the control sample, avoid vigorous
or rapid mixing motion. Visually inspect the control vial to
ensure that the sample is fully reconstituted. If not, discard the
reconstituted fluid and start over with fresh vials.
5
Using a plastic transfer pipette, plastic syringe, or plastic
capillary tube with no anticoagulant, immediately transfer the
solution from the vial into the ACT cartridge
6
Immediately seal the cartridge and insert it into an analyzer.
Note:
Additional ACT cartridges may be tested with the remaining
fluid if used within 30 seconds of complete reconstitution of the
sample.
CONTROLS FOR PT/INR CARTRIDGES
Intended Use
The i-STAT
®
PT Control Level 1 (normal) and PT Control Level 2 (abnormal)
are used to verify the integrity of newly received PT/INR cartridges.
Contents
Each level of control is packaged as a box of 5 vials of lyophilized human
plasma and 5 vials of calcium chloride diluent. The Level 1 control vials contain
9.5
±
1.5 mmol/L calcium chloride, while the Level 2 control vials contain
12
±
2.0 mmol/L calcium chloride.
Storage
i-STAT PT controls, Levels 1 and 2, are contained in 6-mL vials. Separate
6-mL vials contain 1-3 mL of calcium chloride solution for reconstitution.
Refrigerated storage at 2 to 8ºC (35 to 46ºF) should be maintained until the
printed expiration date on the box and vial labels. Do not use beyond the
expiration date on the box and vial labels.
Содержание i-STAT 1
Страница 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Страница 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Страница 44: ......
Страница 90: ......
Страница 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Страница 162: ......
Страница 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Страница 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Страница 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Страница 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Страница 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Страница 192: ......
Страница 194: ......
Страница 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 226: ......
Страница 236: ......
Страница 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Страница 254: ......
Страница 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Страница 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Страница 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Страница 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Страница 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Страница 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Страница 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Страница 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Страница 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Страница 536: ......
Страница 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Страница 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Страница 572: ...Art 715617 00D Rev Date 06 11 08 ...
Страница 578: ...Art 714547 00F Rev Date 06 11 08 ...
Страница 590: ...Art 720735 00B Rev Date 06 11 08 ...
Страница 592: ...Art 721106 00B Rev Date 06 11 08 ...
Страница 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Страница 610: ...Art 720654 00B Rev Date 06 11 08 ...
Страница 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Страница 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Страница 618: ...Art 721296 00A Rev Date 07 16 08 ...
Страница 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Страница 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 658: ......