Kaolin ACT - 2
Art:
715878-00M
Rev. Date: 31-Jan-12
The test is to be used with the i-STAT Portable Clinical Analyzer and the i-STAT 1 Analyzer, but not the
Philips Medical Systems (formerly Agilent Technologies) Blood Analysis Module (BAM).
Contents
Each i-STAT
Kaolin
ACT cartridge provides a sample collection chamber, sensors to detect the coagulation
endpoint, and dry reagents necessary to initiate and allow coagulation. Stabilizers and reagents are coated
on a section of the sensor channel and include the following reactive ingredients:
Reactive Ingredient
Kaolin
Thrombin Substrate
Metrological Traceability
The i-STAT System test for Kaolin Activated Clotting Time measures the time interval required for complete
activation, by kaolin, of the coagulation cascade in arterial or venous whole blood (dimension seconds)
for
in vitro
monitoring of high-level heparin therapy. Presently, no international conventional reference
measurement procedure or international conventional calibrator for
Kaolin
ACT is available.
Kaolin
ACT values
assigned to i-STAT’s controls are traceable to i-STAT’s selected reference measurement procedure, which
employs Celite activated glass reagent tubes, an automated timer and traditional viscometric clot detection
and is run under specified temperature and sample conditions. i-STAT System controls are validated for
use only with the i-STAT System and assigned values may not be commutable with other methods. Further
information regarding metrological traceability is available from Abbott Point of Care Inc..
Expected Values
Test/Abbreviation
Units
Reportable Range
Reference Range
(PREWRM)
Reference Range
(NONWRM)
Activated Clotting
Time/ACT
seconds
50 - 1000*
74 - 137
82- 152
*
The range from 77 - 1000 seconds (PREWRM mode) has been verified through method comparison studies.
Clinical Significance
The ACT is primarily used to monitor a patient’s state of anticoagulation due to heparin that is administered
during a medical or surgical procedure. It is commonly employed in cardiac catheterization, Percutaneous
Transluminal Coronary Angioplasty (PTCA), renal dialysis, hemodialysis, and extra-corporeal circulation
during bypass.
Performance Characteristics
The typical performance data summarized below was collected in health care facilities by health care
professionals trained in the use of the i-STAT System and comparative methods. All data uses the
PREWRM calibration, unless otherwise noted.
Precision data were collected at i-STAT and during clinical trials following a protocol recommended by
i-STAT and using plasma control material. Similar results can be expected in future performance studies
provided the same experimental design and data analysis procedures are followed.
Plasma Control
n
Mean
SD
%CV
Level 1
119
169 seconds
4 seconds
2.0
Level 2
113
409 seconds
21 seconds
5.2
Содержание i-STAT 1
Страница 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Страница 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Страница 44: ......
Страница 90: ......
Страница 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Страница 162: ......
Страница 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Страница 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Страница 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Страница 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Страница 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Страница 192: ......
Страница 194: ......
Страница 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 226: ......
Страница 236: ......
Страница 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Страница 254: ......
Страница 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Страница 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Страница 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Страница 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Страница 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Страница 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Страница 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Страница 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Страница 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Страница 536: ......
Страница 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Страница 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Страница 572: ...Art 715617 00D Rev Date 06 11 08 ...
Страница 578: ...Art 714547 00F Rev Date 06 11 08 ...
Страница 590: ...Art 720735 00B Rev Date 06 11 08 ...
Страница 592: ...Art 721106 00B Rev Date 06 11 08 ...
Страница 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Страница 610: ...Art 720654 00B Rev Date 06 11 08 ...
Страница 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Страница 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Страница 618: ...Art 721296 00A Rev Date 07 16 08 ...
Страница 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Страница 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 658: ......