Art: 714260-00N
Rev. Date: 19-Jan-12
4
Code
Number
Cause/Action
Message on Display
Explanation
42, 43
Cartridge Error
/ Use
Another Cartridge
These codes indicate that the conductometric sensor (code 42)
or the amperometric sensor (code 43) was out of specification.
This could be caused by a pre-burst calibrant pack, dirty
cartridge contact pads, or a dirty connector in the analyzer.
79-81
Cartridge Error
/ Use
Another Cartridge
Bad contact between the thermal probes in the analyzer and
the metalization on the back of the chips in the cartridge trigger
these codes. Causes are: poor metalization of the chips, dirt
on the metalization, or bent or broken thermal probes in the
analyzer.
21
Cartridge Preburst
/ Use
Another Cartridge
This code indicates that the analyzer detected fluid on the
sensors before it should have. Possible causes: mishandling
of cartridges (putting pressure in the center of the cartridge),
poor storage conditions of cartridges (frozen), or rerunning used
cartridges.
31, 34, 44
Unable to Position
Sample
/ Use Another
Cartridge
The analyzer did not detect movement of sample across the
sensors. This could be due to a clot in the sample (especially in
neonates), to not closing the snap closure on the cartridge, or
to an aberrant cartridge.
35, 36
Sample Positioned
Short of Fill Mark
/ Use
Another Cartridge
The cartridge was underfilled. The sample must reach the fill
mark. Try another cartridge.
30, 37
Sample Positioned
Beyond Fill Mark
/ Use
Another Cartridge
The cartridge was overfilled. The sample was past the fill mark.
Try another cartridge.
38, 39
Insufficient Sample
/
Use Another Cartridge
This is most likely due to insufficient sample in the sample
well of the cartridge, but can also be caused by bubbles in the
sample. Try another cartridge and ensure sufficient sample is in
the sample well.
46
Cartridge Error
/ Use
Another Cartridge
The analyzer did not detect movement of sample across the
sensors. This could be due to a clot in the sample (especially in
neonates), to not closing the snap closure on the cartridge, or
to an aberrant cartridge.
47
Cartridge Not Inserted
Properly
/ Reinsert
Cartridge
This code indicates the cartridge or Electronic Simulator
may not be pushed in all the way. Reinsert the cartridge or
Electronic Simulator. If the problem persists and/or the user
is certain the cartridge or Simulator is properly inserted, it
may indicate an analyzer problem. Contact i-STAT Technical
Services or your local support organization for further
assistance.
48
Analyzer Error
/ See
Manual
This code indicates the cartridge or Electronic Simulator may
have been “cocked” when inserted. Push the cartridge or
Simulator straight through the cartridge port. If the problem
persists, and the user is certain the cartridge or Simulator is
properly inserted, it may indicate an analyzer problem. Contact
i-STAT Technical Services or your local support organization for
further assistance.
Содержание i-STAT 1
Страница 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Страница 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Страница 44: ......
Страница 90: ......
Страница 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Страница 162: ......
Страница 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Страница 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Страница 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Страница 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Страница 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Страница 192: ......
Страница 194: ......
Страница 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 226: ......
Страница 236: ......
Страница 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Страница 254: ......
Страница 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Страница 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Страница 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Страница 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Страница 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Страница 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Страница 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Страница 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Страница 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Страница 536: ......
Страница 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Страница 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Страница 572: ...Art 715617 00D Rev Date 06 11 08 ...
Страница 578: ...Art 714547 00F Rev Date 06 11 08 ...
Страница 590: ...Art 720735 00B Rev Date 06 11 08 ...
Страница 592: ...Art 721106 00B Rev Date 06 11 08 ...
Страница 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Страница 610: ...Art 720654 00B Rev Date 06 11 08 ...
Страница 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Страница 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Страница 618: ...Art 721296 00A Rev Date 07 16 08 ...
Страница 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Страница 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 658: ......