ELECTRONIC SIMULATOR
5
Art: 714367-00D
Rev. Date: 25-Aug-11
5-1
The external Electronic Simulator is a stable electronic device, which is inserted
into the cartridge port. The test cycle for the external Electronic Simulator is
about 60 seconds. (The test cycle for the internal simulator is shorter because
it shares the initial part of the test cycle with the cartridge.)
Overview
The Electronic Simulator, external and internal, is a quality control device
for the analyzer’s cartridge signal-reading function. It simulates two levels of
electrical signals that stress the analyzer’s cartridge signal detection function
both below and above measurement ranges.
While the analyzer performs internal electronic checks and calibration during
each test cycle, the Electronic Simulator test provides an independent check
on the ability of the analyzer to take accurate and sensitive measurements of
voltage, current and resistance from the cartridge. An analyzer will pass or
fail this electronic test depending on whether or not it measures these signals
within limits specified in the analyzer software.
The schedule for the Electronice Simulator can be customized to meet local,
state, or national accreditation requirements. A reminder message for the
operator to run the external simulator can be set by the number of hours on
the i-STAT Portable Clinical Analyzer and by the hours or tests on the i-STAT 1
Analyzer. The schedule for the automatic internal Electronic Simulator can be
set by the number of hours on the i-STAT Portable Clinical Analyzer and by the
hours or tests on the i-STAT 1 Analyzer. For details and lockout options, see the
Customization section of this manual.
Note:
All analyzers that pass the Electronic Simulator test are equivalent.
Therefore, any representative number of analyzers that pass the
Simulator test may be used for compliance with regulatory and
accreditation quality assurance procedures. These procedures include
initial performance verification studies, calibration verification,
proficiency testing, and method comparison studies.
Relative Humidity
The Electronic Simulator test will fail if high humidity interferes with the
measurements. Therefore it is not necessary to record humidity where the
analyzers are in use.
Internal Simulator
When the specified time has elapsed since the last Electronic Simulator test
(internal or external), the internal test will automatically be performed when a
cartridge is inserted before the sample is tested, adding about 20 seconds to the
testing cycle.
External Simulator
Содержание i-STAT 1
Страница 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Страница 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Страница 44: ......
Страница 90: ......
Страница 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Страница 162: ......
Страница 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Страница 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Страница 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Страница 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Страница 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Страница 192: ......
Страница 194: ......
Страница 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 226: ......
Страница 236: ......
Страница 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Страница 254: ......
Страница 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Страница 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Страница 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Страница 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Страница 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Страница 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Страница 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Страница 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Страница 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Страница 536: ......
Страница 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Страница 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Страница 572: ...Art 715617 00D Rev Date 06 11 08 ...
Страница 578: ...Art 714547 00F Rev Date 06 11 08 ...
Страница 590: ...Art 720735 00B Rev Date 06 11 08 ...
Страница 592: ...Art 721106 00B Rev Date 06 11 08 ...
Страница 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Страница 610: ...Art 720654 00B Rev Date 06 11 08 ...
Страница 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Страница 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Страница 618: ...Art 721296 00A Rev Date 07 16 08 ...
Страница 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Страница 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 658: ......