Abbott Point of Care Inc.
• Abbott Park, IL 60064 • USA
Art:
730077-00B
Rev. Date: 14-Mar-12
TECHNICAL BULLETIN
i-STAT
®
LIQUID QUALITY CONTROL SCHEDULE AND LOCKOUT
CUSTOMIZATION ON THE i-STAT 1 HANDHELD
OVERVIEW
As part of the on-going READi initiative (
Responds, Enhances, And Delivers Innovation), Abbott Point
of Care (APOC) has released a new Liquid Quality Schedule and Lockout feature for the i-STAT 1
handheld. This new feature allows a system administrator (e.g. Point of Care Coordinator or Laboratory
Administrator) to define a liquid quality control plan for their i-STAT 1 handhelds and enables them to
enforce quality control (QC) compliance by invoking automatic lockout of patient and proficiency testing
pathways if QC requirements in the plan are not fulfilled. By default, the feature is turned off.
This feature can be used in conjunction with the new Liquid QC Pass/Fail customization to automate and
manage liquid quality control testing on the i-STAT 1 handheld. Please see the Technical Bulletin “Liquid
Quality Control Pass/Fail Customization on the i-STAT 1 Handheld” for more information.
Functionality
Previously, system administrators had to “manually” manage and enforce QC requirements for the i-STAT
1 System, since no automated liquid QC lockouts were provided. With this new customization feature,
the system administrator can define a customized QC plan that includes:
•
which cartridge types and liquid QC fluids must be run,
•
which cartridge types are enabled by running the liquid QC, and
•
under what schedule the liquid QC is to be performed.
A. QC Profile
Using the Customization Workspace in the i-STAT CDS application or i-STAT/DE, the system
administrator defines a number of QC
cartridge sets
, consisting of:
•
a
QC cartridge type
(i.e. the cartridge type to be tested with specified liquid QC fluids during the
QC procedure), as well as
•
any number of
dependent cartridge types
(i.e. associated cartridge types that will be enabled
by the analyzer if the QC requirements for a given cartridge set are met on that analyzer. A QC
cartridge set is allowed to have zero dependent cartridges).
Note: A QC cartridge in one QC cartridge set cannot be a dependent cartridge in another QC cartridge
set within the same schedule, but it can be a dependent cartridge in a QC cartridge set belonging to a
different schedule.
The system administrator also creates a
QC test profile
by associating the defined cartridge set with up
to six (6) specific QC fluids (i.e. specified types and levels of liquid control fluid). All APOC control fluids
as well as three generic (user-defined) fluids are available for selection in the QC test profile.
Содержание i-STAT 1
Страница 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Страница 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Страница 44: ......
Страница 90: ......
Страница 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Страница 162: ......
Страница 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Страница 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Страница 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Страница 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Страница 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Страница 192: ......
Страница 194: ......
Страница 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 226: ......
Страница 236: ......
Страница 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Страница 254: ......
Страница 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Страница 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Страница 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Страница 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Страница 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Страница 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Страница 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Страница 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Страница 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Страница 536: ......
Страница 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Страница 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Страница 572: ...Art 715617 00D Rev Date 06 11 08 ...
Страница 578: ...Art 714547 00F Rev Date 06 11 08 ...
Страница 590: ...Art 720735 00B Rev Date 06 11 08 ...
Страница 592: ...Art 721106 00B Rev Date 06 11 08 ...
Страница 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Страница 610: ...Art 720654 00B Rev Date 06 11 08 ...
Страница 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Страница 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Страница 618: ...Art 721296 00A Rev Date 07 16 08 ...
Страница 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Страница 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 658: ......