Rev. Date: 11-Jul-11
Art: 714173-00M
NA - 3
Slope
1.00 0.98 0.95
Int’t
-0.11 3.57 5.26
Sy.x
1.17 1.04 1.53
Xmin
126 120 124
Xmax
148 148 148
r
0.865 0.937 0.838
Cartridge Comparison
The performance characteristics of the sensors are equivalent in all cartridge configurations. System
difference analysis was performed on 40 patient samples using the i-STAT 6+ and i-STAT EC4+
cartridges. In the 130–150 mmol/L range the average difference was 0.750.
Factors Affecting Results*
Sodium heparin may increase sodium results up to 1mmol/L
7
.
Hemodilution of the plasma by more than 20% associated with priming cardiopulmonary bypass
pumps, plasma volume expansion or other fluid administration therapies using certain solutions
may cause clinically significant error on sodium, chloride, ionized calcium and pH results. These
errors are associated with solutions that do not match the ionic characteristics of plasma. To avoid
these errors when hemodiluting by more than 20%, use physiologically balanced multi-electrolyte
solutions containing low-mobility anions (e.g. gluconate) such as Normosol
®
-R (Abbott Laboratories),
Plasma-Lyte
®
-A (Baxter Healthcare Corporation), and Isolyte
®
-S (B Braun Medical) rather than solutions
such as normal saline or Ringer’s Lactate.
Test concentrations used were as per the CLSI guidance document,
8
unless otherwise indicated.
The following substances are known to interfere with the i-STAT sodium assay:
Substance
Test Concentration
(mmol/L)
Interference
Bromide
37.5
Increased i-STAT Sodium results. See
Note below.
The following substances are known not to significantly interfere with the i-STAT sodium assay at
the stated test concentrations:
Substance
Test Concentration
(mmol/L)
Acetaminophen
1.32
Acetylcysteine
10.2
Ascorbate
0.34
Bromide
(therapeutic)
2.5
9,10,11
β-Hydroxybuterate
6.0
12
Lactate
6.6
Magnesium Chloride
1.0
Salicylate
4.34
Note:
1) Bromide has been tested at two levels; the CLSI recommended level and a therapeutic plasma
concentration level of 2.5 mmol/L. The latter is the peak plasma concentration associated with halothane
anesthesia, in which bromide is released. APOC has not identified a therapeutic condition that would
lead to levels consistent with the CLSI recommended level. Bromide at a concentration of 37.5 mmol/L
increased i-STAT sodium results, while a therapeutic range of bromide (2.5 mmol/L) did not significantly
interfere with i-STAT sodium results.
*It is possible that other interfering substances may be encountered. The degree of interference at concentrations other than those listed might not
be predictable.
Содержание i-STAT 1
Страница 1: ...i STAT 1 System Manual Rev Date 20 Jan 12 Art 714336 00L ...
Страница 18: ...1 8 Art 714363 00U Rev Date 02 Aug 12 ...
Страница 44: ......
Страница 90: ......
Страница 106: ...9 12 Art 714371 00E Rev Date 06 Aug 12 ...
Страница 162: ......
Страница 166: ...Art 730292 00A Rev Date 20 Feb 12 4 ...
Страница 176: ...10 10 Art 714372 00L Rev Date 30 Jul 12 ...
Страница 182: ...11 6 Art 714373 00F Rev Date 17 Jan 11 ...
Страница 184: ...12 2 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 185: ...Art 714374 00J Rev Date 25 Aug 11 12 3 ...
Страница 186: ...12 4 Art 714374 00J Rev Date 25 Aug 11 ...
Страница 187: ...Art 714374 00J Rev Date 25 Aug 11 12 5 ...
Страница 192: ......
Страница 194: ......
Страница 214: ...14 20 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 216: ...14 22 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 218: ...14 24 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 220: ...14 26 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 222: ...14 28 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 224: ...14 30 Art 714376 00M Rev Date 31 Jan 12 ...
Страница 226: ......
Страница 236: ......
Страница 252: ...17 4 Art 714379 00F Rev Date 18 Jul 12 ...
Страница 254: ......
Страница 294: ...Art 725768 00E Rev Date 04 Jan 12 20 ...
Страница 336: ...22 2 Art 714384 00D Rev Date 07 18 06 ...
Страница 342: ...22 8 Art 714384 00D Rev Date 07 18 06 ...
Страница 348: ...22 14 Art 714384 00D Rev Date 07 18 06 ...
Страница 352: ...22 18 Art 714384 00D Rev Date 07 18 06 ...
Страница 364: ...22 30 Art 714384 00D Rev Date 07 18 06 ...
Страница 384: ...22 50 Art 714384 00D Rev Date 07 18 06 ...
Страница 404: ...22 70 Art 714384 00D Rev Date 07 18 06 ...
Страница 488: ...Kaolin ACT 4 Art 715878 00M Rev Date 31 Jan 12 ...
Страница 536: ......
Страница 556: ...Art 714261 00E Rev Date 10 Jan 11 8 ...
Страница 560: ...Art 716240 00D Rev Date 03 Mar 10 4 ...
Страница 572: ...Art 715617 00D Rev Date 06 11 08 ...
Страница 578: ...Art 714547 00F Rev Date 06 11 08 ...
Страница 590: ...Art 720735 00B Rev Date 06 11 08 ...
Страница 592: ...Art 721106 00B Rev Date 06 11 08 ...
Страница 598: ...Art 722831 00A Rev Date 19 Feb 09 ...
Страница 610: ...Art 720654 00B Rev Date 06 11 08 ...
Страница 614: ...Art 721317 00D Rev Date 13 Aug 12 4 ...
Страница 616: ...Art 725703 00A Rev Date 08 Mar 10 2 ...
Страница 618: ...Art 721296 00A Rev Date 07 16 08 ...
Страница 620: ...REV DATE 16 Oct 12 ART 714446 00L ...
Страница 627: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 6 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 628: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 7 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 629: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 8 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 630: ...PROCEDURE MANUAL FOR THE i STAT SYSTEM 9 REV DATE 16 Oct 12 ART 714446 00L ...
Страница 658: ......