11
With the FZ-
i
WP/ FX-
i
WP series
The meaning of IP-65 is "No ingress of dust. Projected against water jets".
If a powerful water jet is used or the balance is immersed in water, it may cause a
damage that is due to ingress of water.
Confirm that "the plug is inserted firmly into the jack" and "the terminal is covered using
the waterproof cover or the waterproof RS-232C cable (AX-KO2737-500)", when using
the balance.
If the terminal cover for the RS-232C is removed from the balance or the waterproof
RS-232C cable (AX-KO2737-500) is not used for the balance, the balance is prevented
from having the dustproof and waterproof performance.
Caution:
Do not install the balance where flammable or corrosive gas is present.
3.2.
During Use
Discharge static electricity from the material to be
weighed (hereinafter referred to as the sample). When
a sample could have a static charge, the weight data is
influenced. Ground the balance and try the following:
Eliminate static electricity using the optional AD-1683
DC static eliminator.
Try to keep the ambient humidity above 45%RH.
Use a metal shield case for a charged sample.
Wipe charged plastic samples with a moistened
cloth.
This balance uses a strong magnet as part of the
balance assembly, so please use caution when
weighing magnetic materials such as iron. If there is a
problem, use the underhook on the bottom of the
balance to suspend the material away from the
influence of the magnet.
Eliminate the temperature difference between a sample
and the environment. When a sample is warmer
(cooler) than the ambient temperature, the sample will
be lighter (heavier) than the true weight. This error is
due to a rising (falling) draft around the sample.
Make each weighing gently and quickly to avoid errors caused by the evaporation of moisture
from the sample or the absorption of moisture by the sample.
Do not drop things upon the weighing pan, or place a sample on the pan that is beyond the
balance weighing capacity. Place a sample in the center of the weighing pan.
Do not use a sharp instrument such as a pencil to press the keys. Use your finger only.
Press the RE-ZERO key before each weighing to eliminate possible errors.
Calibrate the balance periodically so as to eliminate possible errors.
Take into consideration the affect of air buoyancy on a sample when more accuracy is required.
Keep the balance interior free of dust and foreign materials.