47
10.5. Description of the Item “Data Output Mode”
The parameter setting of the “Data output mode (
prt
)” applies to the performance when the data is
transmitted using the RS-232C serial interface.
Key mode
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data and the display blinks one time.
Required setting
dout
prt 0
Key mode
Auto print modes A and B
When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference”
and reference value are met, the balance outputs the weight data.
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the data
and the display blinks one time.
Mode A: Example
“For outputting the weight value each time a sample is added, with
“
ar-d
” set to “
1
” (to adjust zero after the data is output).”
Required setting
dout
prt 1
Auto print mode A (reference = zero)
dout
ap-p
Auto print polarity
dout
ap-b
Auto print difference
dout
ar-d 1
Auto re-zero
Mode B: Example
“For outputting the weight value while a sample is added.”
Required setting
dout
prt 2
Auto print mode B (reference = last stable value)
dout
ap-p
Auto print polarity
dout
ap-b
Auto print difference
Stream mode
The balance outputs the weight data continuously regardless of the display condition. The display
does not blink in this mode.
Example
“For monitoring data on a computer”
With the
FZ-GD
/
FX-GD series, always display weighing values using
the AD-8920A remote display.
Required setting
dout
prt 3
Stream mode
ba5fnc
5pd
Display refresh rate
5if
bp5
Baud rate
Caution
The balance may not transmit the data completely at the specified refresh rate and baud rate.
Set the baud rate higher.
Key mode B
When the PRINT key is pressed, the balance outputs the weight data, regardless of the display
condition. The display does not blink in this mode.
Required setting
dout
prt 4
Key mode B (Immediately)
Key mode C
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data. When the stabilization indicator is not on, the balance waits for the indicator to turn on and
outputs the data. The balance blinks one time in this mode.
Required setting
dout
prt 5
Key mode C (When stable)