38
8.6.
Correcting the Internal Mass Value: 2 (Only for FZ-
i
/
FZ-
i
WP
/
FZ-GD series)
The FZ-
i
/
FZ-
i
WP
/
FZ-GD series balance can correct the internal mass value using an external weight.
Method 2: Calibrate the balance using the external weight. The balance performs automatic
calibration using the internal mass, corrects the internal mass value and stores it in
the balance. The corrected mass value is maintained in non-volatile memory even if
the AC adapter is removed.
The usable calibration weights and the adjustable range are as follows:
Model
Usable calibration weight
Adjustable range
FZ-120
i
FZ-120
i
WP FZ-120GD
100
g
,
50
g
-0.050 g to +0.050 g
FZ-200
i
FZ-200
i
WP FZ-200GD
200
g
,
100
g, 50
g
*1
FZ-300
i
FZ-300
i
WP FZ-300GD
300
g, 200
g
,
100
g, 50
g
*1
FZ-500
i
500
g
,
400
g, 300
g, 200
g
100
g, 50
g
FZ-1200
i
FZ-1200
i
WP FZ-1200GD 1000
g
,
500
g
-0.50 g to +0.50 g
FZ-2000
i
FZ-2000
i
WP FZ-2000GD 2000
g
,
1000
g, 500
g
*1
FZ-3000
i
FZ-3000
i
WP FZ-3000GD 3000
g, 2000
g
,
1000
g, 500
g
*1
FZ-5000
i
5000
g
,
4000
g, 3000
g, 2000
g
1000
g, 500
g
The calibration weight in bold type: factory setting
The calibration weight value can be adjusted within the range above.
*1 Can be selected with balance software version 3.10 or later.
Operation
1 With the factory setting, the balance can not correct
the internal mass value. Refer to “9. FUNCTION
SWITCH AND INITIALIZATION” and set the internal
mass value correction switch to
1
.
2 In the weighing mode, press an d hold the SAMPLE
key to display
ba5fnc
.
3 Press the SAMPLE key several times until
C5 in 2
1
is
displayed.
If
C5 in 2
is not displayed, perform step 1.
4 Press the PRINT key to display
Cal 0
.
Refer to “8-5 Calibration Using an External
Weight”, to perform the calibration.
5 After calibration,
end
is displayed. The balance displays
1
ready
?
when the weight is removed.
Place the accessory breeze break on the balance and
press the PRINT key.
6
1
Cal.
SET
is displayed and the balance corrects the
internal mass value automatically
7 After correcting the internal mass value, the balance displays
C
Cal.
, and performs the calibration automatically using
the corrected internal mass value.
8 The balance displays
end
and returns to the weighing mode.
9 Confirm that the internal mass value has been corrected
properly by placing the external weight used for the correction
procedure on the weighing pan.
If the internal mass value has not been corrected properly,
repeat the procedure to adjust the correction value.
200 g