background image

 

60 

13.

  PROGRAMMABLE-UNIT 

This is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary 

coefficient set in the function table and displays the result.  

The coefficient must be within the range between the minimum and maximum shown below. If the 

coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient 

setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory. 

Model 

Minimum 

coefficient 

Maximum 
coefficient 

FX/FZ 
series  120

i

 /200

i

 /300

/500

i

 

120

i

 WP/200

i

 WP /300

WP  

120GD/200GD/300GD 

0.000001 

1000 

FX/FZ 
series  1200

i

 /2000

i

 /3000

/5000

i

 

1200

i

 WP /2000

i

 WP /3000

WP  

1200GD/2000GD/3000GD 

100 

 
Operation 

1  Press and hold the SAMPLE key until  

ba5fnc 

 of the function table is displayed. 

2  Press the SAMPLE key several times to display  

mlt

j

 . 

3  Press the PRINT key. The balance enters the mode to confirm or set the coefficient. 

Confirming the coefficient 

4  The current coefficient is displayed with the first digit blinking.  

 

 

When it is not to be changed, press the CAL key and go to step 6. 

 

When it is to be changed, press the RE-ZERO key and go to step 5. 

Setting the coefficient 

5  Set the coefficient using the following keys. 

SAMPLE key 

To select a  digit  to change  the 

value. The selected digit blinks. 

RE-ZERO key 

To change the value. 

MODE key 

To change the decimal point 

position. 

 

Each time the switch is pressed, 

the decimal point position 

changes as follows: 

 

 

 

PRINT key 

To store the new setting, display  

 

end  

and go to step 6. 

CAL key 

To cancel the new setting and 

go to step 6. 

Quitting the operation 

6  The balance displays  

Unit 

. Press the CAL key to exit the programmable-unit function and 

return to the weighing mode. 

Using the function 

Press the MODE key to select the  programmable-unit (no display on  the unit section). Perform 

weighing as described in “6.1. Basic Operation (Gram Mode)”. After weighing, the balance displays 

the result (weight data in grams x coefficient). 

Confirm

Содержание FZ-i WP Series

Страница 1: ...200i WP FZ 300i WP FZ 1200i WP FZ 2000i WP FZ 3000i WP FX 120i WP FX 200i WP FX 300i WP FX 1200i WP FX 2000i WP FX 3000i WP FZ 120GD FZ 200GD FZ 300GD FZ 1200GD FZ 2000GD FZ 3000GD FX 120GD FX 200GD...

Страница 2: ...ed into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject...

Страница 3: ...D FX GD Series Balances 18 6 4 Counting Mode PCS 20 6 5 Percent Mode 22 6 6 Statistical Calculation Mode 23 6 7 Statistical Calculation Mode Example of Use 28 Adapting to the Environment 7 RESPONSE AD...

Страница 4: ...OGRAMMABLE UNIT 60 14 PASSWORD LOCK FUNCTION 61 RS 232C Serial Interface 16 RS 232C SERIAL INTERFACE 69 17 CONNECTION TO Peripheral EQUIPMENT 70 17 1 Connection to the AD 8127 Multi Printer or AD 8126...

Страница 5: ...its with most of the common units used around the world Standard RS 232C serial interface to communicate with a computer and to output the Good Laboratory Practice GLP Good Manufacturing Practice GMP...

Страница 6: ...ed to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in the U S A Compliance With Directives of CE mark This device feature...

Страница 7: ...pan support for FZ 5000i and FX 5000i 1 2 3 Assemble the breeze break as shown below Follow the numbered sequence 4 Weighing pan Pan support Main unit Display Bubble spirit level Leveling Wheninstalli...

Страница 8: ...in unit cover AC adapter jack AC adapter plug RS 232C serial interface Grounding terminal Main unit rear side AC adapter ID label AC adapter Note Please confirm that the AC adapter type is correct for...

Страница 9: ...e in the bubble spirit level is inside the red circle 3 Connect the AC adapter plug to the AC adapter jack on the rear side of the balance and connect the plug on other end of the cord to an outlet If...

Страница 10: ...alance away from equipment which produces magnetic fields Install the balance in a stable place avoiding vibration and shock Corners of rooms on the first floor are best as they are less prone to vibr...

Страница 11: ...harged sample Wipe charged plastic samples with a moistened cloth This balance uses a strong magnet as part of the balance assembly so please use caution when weighing magnetic materials such as iron...

Страница 12: ...diagram for achieving waterproof performance or on the pan support boss there may be cases where proper performance cannot be obtained from the balance Use the balance after cleaning those If the dia...

Страница 13: ...eighing units stored in the function table Refer to 5 WEIGHING UNITS Enters the response adjustment mode Cancels the operation when performing settings With the FZ i FZ iWP FZ GD series enters the cal...

Страница 14: ...MODE key For details about the units and modes see the table below Name unit mode Abbrev Display Function table Storing mode Conversion factor 1 g Gram g g g 1 g Counting mode Pcs pcs pcs Percent mode...

Страница 15: ...4 13 8 0 00005 Tael Taiwan 3 25 5 86 8 53 13 8 0 00005 Tael China 3 90 7 04 10 2 16 6 0 00005 Tola India 10 4 18 8 27 4 44 5 0 0001 Messghal 26 0 46 9 68 2 110 0 0005 Unit Capacity Minimum display FZ...

Страница 16: ...order to be displayed using the following keys SAMPLE key To sequentially display the units RE ZERO key To specify a unit or mode The stabilization indicator appears when the displayed unit or mode is...

Страница 17: ...ndicator is on pressing the PRINT key will output the weight value using the RS 232C serial interface Note Peripheral equipment that is sold separately such as a printer or a personal computer is requ...

Страница 18: ...to be displayed Read the value While the stabilization indicator is on pressing the PRINT key will output the weight value using the RS 232C serial interface Note Peripheral equipment that is sold sep...

Страница 19: ...tare subtraction operation is performed About measurement range For the balance the range that can be weighed is determined by model The total amount net amount tare quantity up to the maximum display...

Страница 20: ...counting mode Storing a sample unit mass 2 Press the SAMPLE key to enter the sample unit mass storing mode Even in the storing mode pressing the MODE key will switch to the next mode 3 To select the n...

Страница 21: ...s to be counted on the pan While the stabilization indicator is on pressing the PRINT key will output the weight value number of objects using the RS 232C serial interface Note Peripheral equipment th...

Страница 22: ...key to store the reference mass The balance displays 100 00 The decimal point position depends on the reference value Notes If the balance judges that the mass of the sample is too light to be used a...

Страница 23: ...the equation below Standard deviation Standard deviation Coefficient of variation CV Average Maximum value Average Relative error of maximum value Average Minimum value Average Relative error of mini...

Страница 24: ...ge maximum minimum average Standard deviation coefficient of variation Relative error of maximum value relative error of minimum value 7 Press the PRINT key to store the setting 8 Press the CAL key to...

Страница 25: ...weight data and includes the weight data to statistical calculation in the weighing mode Output is not in the data format specified in the function table because of the data number added Outputs the...

Страница 26: ...Statistical items are indicated on the upper left of the display using the following symbols Symbol Statistical item SUM Sum MAX Maximum MIN Minimum R Range Maximum minimum AVE Average SD Standard dev...

Страница 27: ...it from statistical calculation The number of data decreases by 1 when the balance returns to the weighing mode Clearing the statistical data All the statistical data will be deleted and the number o...

Страница 28: ...ct the application function Press the SAMPLE key several times to display 1ap fnc Then press the PRINT key to display 1 apf 0 3 Change the application function parameter to 2 Press the RE ZERO key to...

Страница 29: ...he peripheral output equipment is connected 4 Weigh formula 2 and press the PRINT key The balance displays 0 00 g Storing the weight value of formula 2 The weight value data is output when the periphe...

Страница 30: ...he response adjustment Either FAST MID or SLOW can be selected 3 After a few seconds of inactivity or when the PRINT key is pressed the balance displays eend Then it returns to the weighing mode and d...

Страница 31: ...FUNCTION TABLE Calibration test is available only when GLP output info of Data output dout is set to 1 or 2 When calibrating use the accessory breeze break All models of the FZ i series FZ i WP serie...

Страница 32: ...displays Calin and performs calibration using the internal mass Do not allow vibration or drafts to affect the balance 4 The balance displays end after calibration If the GLP output info parameter of...

Страница 33: ...use the calibration weight value stored in the balance go to step 5 4 Specify the calibration weight value as follows SAMPLE key To switch the display condition to All of weight selection mode or The...

Страница 34: ...ght Do not allow vibration or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays 1 glp and outputs Calibration Rep...

Страница 35: ...weights is shown on page 31 press the SAMPLE key and go to step 4 If you use the target weight value stored in the balance go to step 5 4 Specify the target weight value as follows SAMPLE key To swit...

Страница 36: ...et weight value 6 Place the displayed target weight on the pan and press the PRINT key The balance measures the target weight and displays the measured value Do not allow vibration or drafts to affect...

Страница 37: ...With the factory setting the balance can not correct the internal mass value Refer to 9 FUNCTION SWITCH AND INITIALIZATION and set theinternal mass value correction switch to1 3 In the weighing mode p...

Страница 38: ...ove 1 Can be selected with balance software version 3 10 or later Operation 1 With the factory setting the balance can not correct the internal mass value Refer to 9 FUNCTION SWITCH AND INITIALIZATION...

Страница 39: ...to the function table Calibration using the internal mass One touch calibration 0 To inhibit calibration using the internal mass 1 To permit calibration using the internal mass Calibration using an ex...

Страница 40: ...ing the Balance This function returns the following parameters to factory settings Calibration data Function table The sample unit mass value counting mode 100 reference mass value percent mode Extern...

Страница 41: ...that the parameter displayed is in effect When pressed and held for 2 seconds in the weighing mode enters the function table mode Selects the class or item in the function table mode Changes the para...

Страница 42: ...a p on Auto display ON 0 OFF Turns on the weighing mode display when the AC adapter is connected 1 ON poff Auto display OFF 0 OFF Turns off the display after 10 minutes of inactivity 1 ON 10 minutes r...

Страница 43: ...lue ap b Auto print difference 0 10 digits Difference between reference value and displayed value 1 100 digits 2 1000 digits int Interval time 0 Every measurement Interval time for the interval output...

Страница 44: ...ode is selected Unit Unit Refer to 5 WEIGHING UNITS Iid ID number Refer to 11 ID NUMBER AND GLP REPORT Aap fnc Application apf Application function 0 Normal weighing mode 1 Capacity indicator 2 Statis...

Страница 45: ...e width to regard a weight value as a stable value When the fluctuation per second is less than the parameter the balance displays the stabilization indicator and outputs the data The parameter influe...

Страница 46: ...rc 0 The tracking function is not used Used for weighing a very light sample trc 1 The tracking function is used Normal zero tracking trc 2 The tracking function is used Strong zero tracking trc 3 The...

Страница 47: ...utputting the weight value while a sample is added Required setting dout prt 2 Auto print mode B reference laststablevalue dout ap p Auto print polarity dout ap b Auto print difference Stream mode The...

Страница 48: ...s excluding the terminator A header of two characters indicates the balance condition The polarity sign is placed before the data with the leading zeros If the data is zero the plus sign is used The u...

Страница 49: ...table value MT format 5if type 3 This format is used when the peripheral equipment of other manufacturer is connected Please note that the connection is not guaranteed A header of two characters indic...

Страница 50: ...R LF CR LF CR LF CR LF CR LF A D DP KF MT NU E H Unstable LF CR LF CR LF CR LF CR A D DP KF MT NU CR LF CR LF LF Negative error Overload CR LF CR CR A D DP KF MT NU L LF U S 0 0 1 8 3 9 6 g U S 1 8 3...

Страница 51: ...pacity indicator apf 1 In the weighing mode the indicator displays the weight data relative to the weighing capacity in percentage Zero 0 maximum capacity 100 Statistical calculation mode apf 2 The mo...

Страница 52: ...correct and the operation is to be finished press the CAL key and go to step 8 When the date is correct and the time is to be confirmed press the SAMPLE key and go to step 6 When the date is not corr...

Страница 53: ...dar function and return to the weighing mode Notes Do not enter invalid values such as a non existing date when setting the time and date 10 10 Comparator Function The results of the comparison are in...

Страница 54: ...digit selected MODE key To switch the polarity PRINT key To store the new setting and go to step 7 CAL key To cancel the new setting and go to step 7 Entering the lower limit value 7 With Cp lo displa...

Страница 55: ...utput info parameter to 1 The balance can output the following Calibration report of the calibration using the internal mass Calibration due to changes in temperature and one touch calibration Calibra...

Страница 56: ...e AD 8127 Multi Printer To output the report to a personal computer using the RS 232C serial interface set the GLP output info parameter to 2 Note If the time and date are not correct adjust the AD 81...

Страница 57: ...fo 2 Calibration test report usingan external weight Printoutexamples for the FX i FX iWP FX GD series Calibration test does not perform calibration When the setting is info 1 When the setting is info...

Страница 58: ...i FX i WP FX GD series 1 With the weight data displayed press and hold the PRINT key until S5tart is displayed The Title block is output 2 The weight data is output according to the parameter setting...

Страница 59: ...is revealed by removing the plastic cap on the bottom of the balance Use the underhook as shown below Caution Do not apply excessive force to the underhook When not in use attach the plastic cap to p...

Страница 60: ...the mode to confirm or set the coefficient Confirming the coefficient 4 The current coefficient is displayed with the first digit blinking When it is not to be changed press the CAL key and go to ste...

Страница 61: ...d setting changes are also possible pa55 1 Request password input at the start of weighing An administrator admIN can limit the users of the balance by setting individual passwords The password input...

Страница 62: ...tting of date and time 6 The administrator admIN can set this to prohibited as shown in 9 FUNCTION SWITCH AND INITIALIZATION Login level Weighing Password input at weighing start Calibration Change th...

Страница 63: ...key to display 5urE No YES No blinking while No selected 6 Press the RE ZERO key to switch No YES 7 Display 5urE No YES YES blinking when selected YES 8 Press the PRINT key while YES is selected to en...

Страница 64: ...after all lamps are illuminated After entering the administrator s password log in as an administrator At factory settings the password is set to 2222 by RE ZERO key input 4 times at the administrator...

Страница 65: ...will be displayed after all lamps are displayed At factory settings the password is set with 2222 of RE ZERO key input 4 times at the administrator level If the password is incorrect the buzzer sounds...

Страница 66: ...ey to change the login level admIN U5Er 01 to 10 If the password is already registered at the login level the stability mark is lit changeable 7 Press the PRINT key to change the password Refer to Cha...

Страница 67: ...is selected Press the CAL key to return to the 4th character input 6 Press the RE ZERO key to display 5urE No YES where YES is blinking when YES is selected 7 Press the PRINT key to store the new pas...

Страница 68: ...and Changing Password to change an arbitrary password Missing Password If the correct password is missing the balance can not be used Contact your local A D dealer to reset the password to factory set...

Страница 69: ...Signal name Direction Description 1 N C samepotential withSG 1 2 TXD Output Transmit data 3 RXD Input Receive data 4 No connection 5 SG Signal ground 6 DSR Output Data set ready 7 RTS Input Request to...

Страница 70: ...output No setting DUMP Refer to the instruction manual of the AD 8127 for how to change internal settings of the AD 8127 multi printer Internal settings of the balance Usage example Balance prt Data...

Страница 71: ...t appropriate parameters for ap p Auto print polarity and ap b Auto print difference when auto print mode A or B prt 1 or 2 is selected Set the AD 8121B DIP switch No 3 to ON when unstable data is pri...

Страница 72: ...Can make bi directional communication between the balance and a personal computer using the RS 232C serial interface Can display or store the data using a text file format Can also print the data usi...

Страница 73: ...he balance GLP report can be printed using a printer connected to a personal computer 4 Receiving weight data at a certain interval The weight data can be received at a certain interval and data chara...

Страница 74: ...display off ON Turns the display on P Same as the ON OFF key PRT Same as the PRINT key R Same as the RE ZERO key SMP Same as the SAMPLE key T Tare key Z Same as the RE ZERO key E SCT Same as the RE Z...

Страница 75: ...to control the balance and can process it the balance transmits the acknowledge code Among commands to control the balance the following transmit the acknowledge code both when the balance receives t...

Страница 76: ...mation of command receipt Confirmation of command receipt Command completion Starts inputting data Zero display Calibration weight value display Starts inputting data RE ZERO operation L CR LF A C AKC...

Страница 77: ...ance has two functions Data output dout and Serial interface 5if Set each function as necessary PT command Tare of 100g PT command LF CR T P PT 00100 00 gCR LF P T 1 0 0 Tare value display AKCR LF g C...

Страница 78: ...ing material for transportation Specifications of the FZ iWP FX iWP series In normal use the balance can be cleaned with water But keep the following precautions so that dust and water do not invade t...

Страница 79: ...f 0 001 g Is the balance level Refer to 3 1 Before Use Is the operating environment free from vibration and drafts For the minimum weighing value of 0 001 g has the breeze break been installed Is ther...

Страница 80: ...ing mode EC E21 Calibration weight error The calibration weight is too light Confirm that the weighing pan is properly installed Confirm the calibration weight value Press the CAL key to return to the...

Страница 81: ...ed a Q command but not in the weighing mode e g The balance received a Q command while processing a RE ZERO command Adjust the delay time to transmit a command EC E03 Timeout error If the timeout para...

Страница 82: ...put statistical data or GLP data to a personal computer use the USB converter AX USB 9P EX FXi 08 Ethernet interface Used to connect the balance to a LAN The WinCT Plus data communication software is...

Страница 83: ...standard accessory AD 1654 Density Determination Kit Example Temperature Water density 0 C 0 99984 g cm3 10 C 0 99970 g cm3 20 C 0 99820 g cm3 30 C 0 99565 g cm3 Models the FXi 10 small breeze break...

Страница 84: ...possible to store data in an environment where a computer can not be used The stored data can be read to a personal computer using USB As the AD 1687 is recognized as USB memory special software is n...

Страница 85: ...onditioning or people passing by thereby reducing balance weighing error The transparent panel assembly consists of antistatic plastic material that protects the balance from static electricity AD 168...

Страница 86: ...ector is used for the RS 232C interface of the balance AD 8529PC W Bluetooth converter for connecting a personal computer This option connects a personal computer to the balance wirelessly via Bluetoo...

Страница 87: ...0 g Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored External calibration weight 100 g 50 g 200 g 100 g 50 g 300 g 200 g 100 g 50 g 500 g 400 g 300 g 200 g 100 g 50 g 1000 g 500 g 2...

Страница 88: ...nd or 20 times second Counting mode Minimum unit mass 0 001 g 0 01 g Number of samples 5 10 25 50 or 100 pieces Percent mode Minimum 100 reference mass 0 100 g 1 00 g Minimum 100 display 0 01 0 1 1 De...

Страница 89: ...00i WP FX 120GD 200GD 300GD FZ 1200i 2000i 3000i 5000i FZ 1200i WP 2000i WP 3000i WP FZ 1200GD 2000GD 3000GD FX 1200i 2000i 3000i 5000i FX 1200i WP 2000i WP 3000i WP FX 1200GD 2000GD 3000GD Unit mm Th...

Страница 90: ...rational function Re zero To set the display to zero GLP Good Laboratory Practice Repeatability Variation in measured values obtained when the same weight is placed and removed repetitively Usually ex...

Страница 91: ...ap b Auto print difference 43 71 apf Application function 23 44 ap p Auto print polarity 43 71 Application 23 44 Application function 23 44 ar d Zero after output 43 at f Auto feed 43 Auto display OF...

Страница 92: ...9 72 Decimal point 42 46 Density Determination Kit 83 Diaphragm for waterproof 7 Digit 42 46 90 Display at start 42 Display refresh rate 42 46 87 88 dout Data output 43 DP format 44 48 DTE 69 Dustproo...

Страница 93: ...hing value 87 88 mlt Programmable unit 44 MODE key 13 15 16 18 MT format 44 49 NU format 44 49 Number of data 23 44 OK buzzer 42 ON OFF key 13 40 Operating environment 87 88 pa55 44 61 Pan support 7 P...

Страница 94: ...Date output 43 Stop bit 69 Stream mode 43 47 SUM Sum 26 Sum 23 26 44 Tare 17 18 20 22 29 Terminal cover 8 78 Terminator 44 Time 50 Time Date output 43 Timeout 45 Title block 58 Transmission form 69 Tr...

Страница 95: ......

Страница 96: ...w Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 6 33 817 07331 817 Manhattan Bldg 33 Gukjegeumyung ro 6 gil Yeongdeungpo gu Seoul 07331...

Отзывы: