21
Notes
If the balance judges that the mass of the samples is too light and is not adequate to be used
as the unit mass, it displays
lo
. In that case, store the mass by some quantity. For example,
when the model with the minimum weighing value of 0.01 g is used and 10 pieces of samples
weigh 0.05 g. Store 100 pieces of samples as 10 and multiply the weighing result by 10.
If the balance judges that the mass of the samples is too light to acquire accurate weighing, it
displays an error requiring the addition of more samples to the specified number. For example,
1
50-
PCS
appears to require 25 more samples. Add 25 samples and press the PRINT key.
When the unit mass is stored correctly, the balance proceeds to the counting mode.
For a more accurate counting operation, use samples with a unit mass of at least ten or more
times the minimum weighing value of the balance. For example, when a balance with a
minimum weighing value of 0.01 g is used, the recommended sample unit mass is 0.1 g or
more.
The sample unit mass stored, even if the AC adapter is removed, is maintained in non-volatile
memory.
Counting operation
7 Place the samples to be counted on the pan.
While the stabilization indicator is on, pressing the PRINT
key will output the weight value (number of objects), using
the RS-232C serial interface.
Note
Peripheral equipment, that is sold separately , such as a
printer or a personal computer is required.
Counting mode using the ACAI function
The ACAI is a function that improves the accuracy of the unit
mass automatically by increasing the number of samples and
averaging the unit mass variable to minimize the weighing
error, as the counting process proceeds.
8 If a few more samples are added, the processing indicator turns on. To prevent an error, add
three or more. The processing indicator does not turn on if overloaded. Try to add the same
number of samples as displayed.
9 The balance re-calculates the unit mass while the processing indicator is blinking. Do not touch
the balance or samples on the pan until the processing indicator turns off.
10 Counting accuracy is improved when the processing indicator turns off.
Each time the above operation is performed, a more accurate unit mass will be obtained. There
is no definite upper limit of ACAI range for the number of samples exceeding 100. Try to add
the same number of samples as displayed.
11 Remove all the samples used in ACAI and proceed with the counting operation using the
improved unit mass.
Содержание FZ-i WP Series
Страница 95: ......