45
Class
Description
C5 in 1
*
Internal mass value correction: Method 1
Refer
to “8. CALIBRATION”.
Displayed only when the
function switch (internal mass
value correction switch) is set
to
1.
C5 in 2
*
Internal mass value correction: Method 2
*
: Only for the FZ-
i
/
FZ-
i
WP
/
FZ-GD series : Factory settings.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending on
the baud rate.
When a refresh rate of 20 times/second is used, set the baud rate to 4800 bps or higher.
10.4. Description of the Class “Environment, Display”
Condition (
Cond
)
Cond 0
This parameter is for sensitive response to the fluctuation of a weight value.
Used for powder target weighing, weighing a very light sample or when quick
response weighing is required.
After setting, the balance displays FAST.
Cond 2
This parameter is for stable weighing with slow response. Used to prevent a
weight value from drifting due to vibration or drafts.
After setting, the balance displays SLOW.
Note
With “Hold function (
Hold
)” set to “ON (
1
)”, this item is used to set the
averaging time.
Stability band width (
5t-b
)
This item controls the width to regard a weight value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs the data.
The parameter influences the “Auto print mode”
5t-b 0
This parameter is for sensitive response of the stabilization indicator. Used for
exact weighing.
5t-b 2
This parameter ignores slight fluctuation of a weight value. Used to prevent a
weight value from drifting due to vibration or drafts.
Note
With “Hold function (
Hold
)” set to “ON (
1
)”, this item is used to set the
stabilization range.
Hold function (
Hold
) (Animal weighing mode)
This function is used to weigh a moving object such as an animal.
When the weight data is over the weighing range from zero and the display fluctuation is within the
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the animal. When the animal is removed from the weighing
pan, the display returns to zero automatically.
This function is available only when the hold function parameter is set to “
1
”
(the animal mode indicator
ANIMAL illuminates) and any weighing unit other than the counting mode is selected.
The stabilization range and averaging time are set in “Condition (
Cond
)” and “Stability band width (
5t-b
)”.