5 Troubleshooting and Maintenance
•
Imaging error (for example, chromatic aberrations)
•
Contact Thermo Fisher Scientific Service for support.
Warning!
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.
Virtual slides show poor stitching in Pannoramic Viewer
•
Scan camera not properly adjusted (for example, due to loose fixture)
•
Contact Thermo Fisher Scientific Service for support.
Slide is often positioned incorrectly
•
Guiding fault in transport system
•
Contact Thermo Fisher Scientific Service for support.
Digitization process is extremely slow
•
Extended focusing
option is activated
•
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated
•
Deactivate the
Z-stack
option if not required.
Image is out of focus
•
Focus range limit is activated
•
Deactivate the
Limiting focus range
option.
•
Focus range setting is incorrect
•
Change focus range setting.
All individual images show a shading effect
•
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide (see section
4.2.2 Image
Compensation
).
June 4, 2018 - Rev. 2
3DHISTECH Ltd.