Pannoramic DESK II 1.23 User’s Guide
4 Software Description
Brightfield ICC Profile
Select an applicable preset color profile to be used for Brightfield scanning from the drop-down
menu. Settings according to the profile has effect on the color set, color temperature, and gamma
value of the scanned image.
•
Preset profiles
are the following:
◦
Calibrated Linear
: Color scheme is defined by calibration (Gamma value is not included)
◦
sRGB standard
: Color scheme is defined by calibration (standard Gamma value -2.2- is
applied)
◦
sRGB standard Microscope Like
: Color scheme is defined by calibration (standard
Gamma value / White Point: D50)
◦
Legacy
: a Color scheme calibrated for the previous software version is applied
◦
Custom
: parameters are freely customizable as follows:
•
Color Scheme
◦
Calibrated
: Scheme is defined by the color calibration slide
◦
Legacy
: a Color scheme calibrated for the previous software version is applied
◦
Raw
: the color information of the recorded raw image is applied
•
White Point
◦
D65
: color temperature according to a standard illuminant D65 white point value is
applied (6500K / daylight)
◦
D50
: color temperature according to a standard illuminant D50 white point value is
applied (5000K / horizontal daylight)
•
Gamma
◦
Standard
: a value of 2.2 is used
◦
Medium
: a value of 1.8 is used
◦
Off
: a value of 1.0 is used
Save settings by clicking the button.