Pannoramic DESK II 1.23 User’s Guide
5 Troubleshooting and Maintenance
5 Troubleshooting and Maintenance
5.1 Safety Information
Check cables, plugs and connections before operation, and if defective or cannot be replaced, contact
Thermo Fisher Scientific Service for support.
Check whether the device is compatible with your local line voltage. It can operate at line voltages
100V to 240V and 50/60 Hz.
Only plug the device in a socket with protective grounding (earth) contact.
Do not operate the device in explosive environment.
Disconnect the device from the power outlet before opening it or replacing a part.
The product is not equipped with protection against hazardous materials (infectious, toxic,
radioactive, corrosive substances or other effects damaging to health). Decontaminate infectious
materials after use.
Slides that are not used any longer do not classify as domestic waste and must be treated as
hazardous materials and disposed in compliance with the currently valid legal requirements.
Dust, specks and humidity may decrease or harm the operating capability of the device. Protect it
with dust cover when not in use. Switch the power off before covering it.
Avoid touching the lamp bulb. Wait for about 15 minutes to allow the bulb to cool down and
disconnect the power plug before changing the bulbs.
Defective devices do not classify as domestic waste. Dispose of the device in compliance with the
currently valid legal requirements.
5.1.1 Cleaning the device
Before cleaning, switch off the device and pull the plug out.
Do not let liquid penetrate into the system.
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with water.
To clean greasy surfaces, it is recommended to apply
soapy water
on a piece of lint-free cotton cloth
and gently rub the surface.
5.1.2 Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage temperature
above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C with a relative humidity of less than 75%.
June 4, 2018 - Rev. 2
3DHISTECH Ltd.
67
(