Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
ALARM TESTING PROCEDURE (Continued)
No-Clamp Alarm Test
1.
Load an administration set with no air bubble into the
pump. Make sure the IV set based clamp was NOT
installed in pump. Close the pump door.
2.
Program the primary flow RATE at 500 ml/h, and the
primary VTBI at 100 ml.
3.
Press the RUN/STOP key to start the infusion.
4.
The infusion status indicator will turn blinking red,
audio alarm will sound, and the Main Display will
show “NO CLAMP”. The infusion will stop.
5.
Press the RUN/STOP key again to acknowledge the
alarm condition. The Z-800F Pump will be in PAUSE
state. The “PAUSE” message will appear on the
display screen.
6.
Install the IV set based clamp in pump. Press the
RUN/STOP key again to resume infusion.
Battery Test
1.
Connect the Z-800F pump to an approved AC power
outlet for at least 18 hours to allow the battery to fully
charge.
2.
Unplug the power cord from the AC power. Turn on
the device. Verify that the battery indicator
is
ON.
3.
Set a primary infusion with the following:
Pri RATE = 125 ml/h
Pri VTBI = 1000ml/h
4.
Start the infusion. Record the infusion starting time.
5.
Record the time when the Low Battery Warning is
presented.