INFUSION MODE INFORMATION
30
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
TPN MODE INFUSION (Continued)
Programming TPN Mode Infusion
TPN infusion programming screen contains the following
display areas:
1.
Status Bar
: The Status Bar displays current operating
mode, current pump state, and volume infused.
TPN
: TPN mode operation
VINF
: Volume Infused for the current infusion
2.
Infusion Parameters
: TPN total VTBI and Total TIME
of the infusion.
TIME
: Total TIME of the infusion.
VTBI
: Total Volume to Be Infused.
3.
Keys
: Allow user to access pump configuration and TPN
ramp programming screen
CONFIG
: Accesses pump configuration
RAMP
: Accesses TPN Ramp Up and Ramp Down
Time parameter programming
Up/Down Arrows
: Highlight the selected parameter
item
WARNING: Do not over-program VTBI.
Program the actual amount of the fluid in the IV bag.
The TPN infusion parameters may be programmed by using
the UP/DOWN Arrow keys and Data Entry keys.
Access TPN Ramp Time Parameter Programming
From TPN programming screen, press RAMP Key to access
TPN Ramp Parameter programming screen.
Programming TPN Ramp Time Parameters
The TPN Ramp Time parameters may be programmed by
using the UP/DOWN Arrow keys and Data Entry keys.