MAINTENANCE
Z-800F Instructions for Use.
75
P/N 800F-IFU-2602, Rev. O
CLEANING
DO NOT
spray cleaning fluids directly onto the
instrument or immerse the instrument in fluids.
DO NOT
use solutions containing phosphoric acid
(Foamy Q&A*), aromatic solvents (naphtha,
paint thinner, etc.), chlorinated solvents*
(Trichloroethane, MEK, Tuluene, etc.),
ammonia, acetone, benzene, xylene or alcohol,
other than as specified below.
DO NOT
use hard or pointed objects to clean any part of
the instrument.
Acceptable cleaning solutions are:
Warm water
Mild detergent (e.g., Manu-Klenz®)
** 10% bleach solution (1 part bleach to 9 parts water)
** Compublend™ II
** Envirocide®
** 2% Glutaraldehyde in water
** Hydrogen Peroxide 3%
** 70% Isopropyl Alcohol
** 2% Phenols in water (O-Syl 1:128, Pheno-Cen 1:256,
Vesphene®
** 10% Providone lodine (betadine™)
** Quaternaries 1:512
WEX-CIDE
1.
Keep the instrument upright and do not allow any part of
the instrument to become saturated with or submersed in
fluid during the cleaning operation.
2.
Use soft cloth dampened with warm water and a mild
nonabrasive cleaning solution to clean all exposed
surfaces. Do not spray any fluids directly on the
instrument. For sanitizing or antibacterial treatment, use
10% bleach solution and water.
WARNING:
Turn the
instrument off and unplug
the power cord from the AC
power source before
cleaning.
Do not spray fluid directly
onto the instrument.
Do not steam autoclave, EtO
sterilize, immerse the
instrument or allow fluid to
enter the instrument case.
Failure to follow these
instructions may result in an
electrical hazard, damage to
the instrument, and voided
warranty coverage.
CAUTION:
The
solutions/solvents identified
as NOT to be used can
damage the surface of the
instrument.
** After application, rinse all
surfaces with a water-
dampened soft cloth.