ALARMS AND TROUBLESHOOTING
Z-800F Instructions for Use.
47
P/N 800F-IFU-2602, Rev. O
ALARMS AND TROUBLESHOOTING
To enhance safety and ease of operation, the Z-800F Infusion Pump provides a full range
of audio and visual alarms, warnings, and prompts.
DEFINITIONS
Error
An audio and visual signal indicates a failure has been detected.
Immediate action is required.
The affected Z-800F pump needs to be replaced with an operational
unit. The affected pump should be serviced by qualified personnel.
Alarm
An audio and visual signal indicates that a potentially unsafe
condition is detected. Immediate action is required.
Under an alarm condition, the Z-800F Infusion Pump is in STOP
state. The audio signal will sound until positive confirmation from
user is delivered.
The Z-800F Infusion Pump will not allow the user to resume the
infusion until the potentially unsafe condition is resolved.
Warning
An audio and visual signal indicates that a potentially unsafe
condition is present. Immediate action is required.
Under a warning condition, the Z-800F Infusion Pump will continue
to operate. The audio signal will not be silenced until the warning
condition is resolved.