INFUSION MODE INFORMATION
Z-800F Instructions for Use.
35
P/N 800F-IFU-2602, Rev. O
INTERMITTENT MODE INFUSION (Continued)
Programming INTERMITTENT Mode Infusion
INTERMITTENT infusion programming screen contains the
following display areas:
1.
Status Bar
: The Status Bar displays current operating
mode, current pump state, and volume infused.
INTR
: INTERMITTENT mode operation
VINF
: Volume Infused for the current infusion
2.
Infusion Parameters
: INTERMITTENT total VTBI and
Cycle TIME of the infusion.
Total VTBI
: Total Volume to Be Infused
Cycle Time
: This displays the time for each cycle in
Hour
:
Minute format. Cycle time should be
programmed between 30 minutes and 24 hours. If the
programmed Cycle Time is outside the range, pump
alarms “Invalid Parameters.”
3.
Keys
: Allow user to access pump configuration and
INTERMITTENT ramp programming screen.
CONFIG
: Accesses pump configuration
PERIOD
: Accesses INTERMITTENT period
parameter programming
Up/Down Arrows
: Highlight the selected parameter
item
The Total VTBI and Cycle Time parameters may be
programmed by using the UP/DOWN Arrow keys and Data
Entry keys.
Note:
During idle period, the pump runs at KVO rate. Volume
infused during idle period is counted. The pump alarms
“Infusion Completed” once VINF (volume infused) equals
programmed Total VTBI.