INFUSION MODE INFORMATION
20
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
CONTINUOUS MODE-PRIMARY INFUSION (Continued)
Primary Infusion Execution
During an infusion, the two Infusion Status Indicator LEDs
will illuminate in green and toggle once per second.
The primary infusion execution display contains the
following contents:
Status Bar:
The Status Bar displays the current operating
mode, current pump state, and volume infused.
PRI
: Primary infusion
RUN
: Pump RUN state
TL
: Time Left in HH:MM format for the current
infusion
Infusion Parameters:
Current infusion parameters.
RATE field displays primary flow rate.
VTBI field counts down to show remaining VTBI.
VINF field counts up to show volume infused.
3 minutes before infusion completes, a short audio prompt
tone will sound twice. Along with the audio prompt, the
screen will display “INFUSION NEAR END” message.
This alert will repeat every 5 seconds until the infusion is
complete.
At completion of the infusion, an audio prompt sounds and
the screen will display “INFUSION COMPLETE – KVO.”
This alert will repeat every 5 seconds until user
intervention. During KVO state, the Flow Rate Indicator
will change to display KVO flow rate.
Clearing the Volume Infused During a Primary Infusion:
Press the RUN/STOP key to PAUSE the primary infusion. Exit
the PAUSE screen by pressing the PROG key. From the
Primary SETUP
screen, press and hold the
CLR
key. An
audio/visual prompt will be presented requesting user
confirmation to clear VINF. Press the
YES
key to confirm. The
VINF field will be reset to 0 ml. The user can press the
NO
key
to abort the action.
NOTE
: The VINF field will record cumulative infusion
volume infused unless cleared by the user pressing the
CLR
key or turning the pump off/on and starting a new infusion.