GETTING STARTED
8
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
GETTING STARTED
WARNINGS & CAUTIONS
The following Warnings and Cautions should be strictly followed to avoid harm to patients
and pump operators:
WARNING:
Zyno Medical assumes no responsibility for incidents that may occur
if its product is not used in accordance with its product labeling.
WARNING:
Only use Zyno Medical provided IV Sets with the Z-800F Infusion
Pump.
WARNING:
The
Z-800F Infusion Pump operation is strictly limited to trained
operators whose competency in safe Z-800F Infusion Pump operation and safe IV
therapy practices has been tested and proven.
WARNING:
Make sure the pump is stable by fastening securely to an IV Pole, or
resting on a flat surface. IV Poles vary in quality and stability. Avoid fastening the
pump too high on the pole, and test for stability before using.
WARNING:
Prior to use, always verify the proper function of the display, audible
and visual alarms.
WARNING:
Verify there are no kinks in the tubing when loading into the pump.
WARNING:
Always read and follow the instructions that accompany the fluid
container and IV administration sets used. Carefully follow the instructions in this
document for loading, removing, and reloading the IV Set. Adjust the pumping
section of the IV Set every 24 hours, and replace the IV Set within 72 hours.
WARNING:
To prevent free flow, make sure to load the IV Set in the Infusion
Pump before connecting the IV set to a patient.
WARNING:
To prevent free flow, make sure to disconnect the IV Set from the
patient before unloading the IV set from the Z-800F Infusion Pump.
WARNING:
Disconnect the IV Set from the patient before purging air bubbles out
of IV Tubing.
WARNING:
Make sure to close the roller clamp before removing IV administration
set from the Z-800F Infusion Pump.
WARNING:
Do not over-program the VTBI. Program the actual amount of the
fluid in the IV bag.
WARNING:
After starting an infusion, make sure drops are falling in the Drip
Chamber. If no drops are falling, make sure the Roller Clamp is open. If the Roller
Clamp is open and still no drops are falling, replace and dispose of the IV
Administration Set.