INFUSION MODE INFORMATION
32
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
10-STEP MODE RATE/VOLUME INFUSION (Continued)
Programming 10-STEP Mode Rate/Volume Infusion
Similar to the Continuous Mode infusion programming
screen, the 10-STEP R/V infusion programming screen
contains the following display areas:
1.
Status Bar
: The Status Bar displays current
operating mode, current pump state, and volume
infused.
STEP01
: 10-STEP mode operation-Step01
VINF
: Volume Infused for the current
infusion
2.
Infusion Parameters
: Step01 Flow Rate and
Step01 Volume to Be Infused
RATE
: Current Step Flow Rate
VTBI
: Current Step VTBI
3.
Keys
: Allow the user to access pump
configuration and option setup screen.
CONFIG
: Accesses pump configuration
Up/Down Arrows
: Select parameter items and
navigate between current step and
previous/next step parameter programming
Navigate to Previous/Next Step
Use Up/Down Arrow keys to select current step
parameters as well as change to previous or next step
programming screen.
Similar to the Continuous Mode infusion programming
screen, the 10-STEP R/V infusion parameters may be
programmed by using the UP/DOWN Arrow keys and
Data Entry keys. Simply continue to scroll down to
program each step. The ten steps are a scrollable list that
can be edited anytime during programming. Scroll and
carefully review the steps you programmed before
pressing
RUN/STOP
to begin the infusion. Pressing
RUN/STOP
to begin the infusion
locks in the list. To edit
the list after an infusion is underway, press
RUN/STOP
to
pause
,
and then press the
PROG
key to edit the list.
WARNING: Do not over-
program VTBI. Program the
actual amount of the fluid in the
IV bag.