MANAGING INFUSION
Z-800F Instructions for Use.
39
P/N 800F-IFU-2602, Rev. O
MANAGING INFUSION
PRIME THE ADMINISTRATION SET USING PRIME KEY
The user can use the
PRIME
key to prime air bubbles out
of the IV Administration Set from the distal end after it is
loaded in the pump.
1.
Press the
PRIME
key. The pump screen will display
the PRIME warning message “Disconnect patient,
then press and hold soft “Prime” key as indicated
below.” Confirm that the patient is not connected to
the administration set. Press and hold the
key to
prime.
2.
During the prime, the pump will run at the fixed
prime flow rate of 600mL/hr. Both Air-In-Line
sensor and Occlusion sensor are disabled.
3.
Release the
key when the set is fully primed.
While the
key is pressed, the pump screen will
report the volume primed. The maximum prime
volume is 10mL. Upon reaching the maximum prime
volume, the prime will stop.
WARNING: Patient must be disconnected before
utilizing the PRIME key.