MAINTENANCE
76
Z-800F Instructions for Use
P/N 800F-IFU-2602, Rev. O
INSPECTION REQUIREMENTS
To ensure the system remains in good operating condition,
both regular and periodic inspections are required.
REGULAR INSPECTIONS:
Regular inspection consists of a visual inspection for
damage and cleanliness, and performing the procedure
described in the Start-Up Sequence section of this
instruction for use before each usage of the instrument.
Regular inspections are not covered under the contract or
agreement offered by Zyno Medical and must be
performed by the user.
REGULAR INSPECTIONS
PROCEDURE FREQUENCY
INSPECTION FOR DAMAGE:
Enclosure Each usage
Power Cord Each usage
CLEANING As Required
START-UP SEQUENCE Each usage
NOTE:
If the instrument does not pass the regular
inspection, the affected instrument must be removed from
use and inspected by qualified service personnel.
PERIODIC PREVENTIVE MAINTENANCE:
The pump has a built-in maintenance cycle odometer. The
odometer indicates how many liters of fluid can be
accurately delivered before the pump must be serviced.
This feature is designed to alert users for the required
servicing timing for high-usage pumps.
When the value of the odometer approaches zero, the pump
will display a warning message after power on, “Pump will
be due for service. Please call 1-866-395-1988.”
When the value of the odometer becomes zero or negative,
the pump will display a warning message after power on,
“Pump is due for service. Please call 1-866-395-1988.”
WARNING:
Failure
to perform these inspections
and maintenance may result
in improper instrument
operation.