X-Ray Inspection System
User Manual
P a g e
| 62
Copyright©2021 ZKTECO CO., LTD. All rights reserved.
5.3.3
Inspection of Lead Shield at the Entrance and Exit of the
Passage
If the interval spaces of the lead shield are large or it falls off and gets damaged, please
contact the technician for consultation and repair.
5.3.4
Inspection of X-Ray Light Indicator and Power Indicator
Light
During the operation of the equipment, if the X-Ray indicator light (red) or power
indicator light cannot light up normally, please contact the technician for consultation
and repair.
5.3.5
Inspection of Emergency Stop Button
The emergency stop buttons at the entrance and exit of the equipment and on the
special-purpose keyboard are the critical safety buttons. When the equipment is
inspecting the articles normally, and if the emergency stop button is pressed, the
conveyor stops running immediately and X-Ray indicator light goes out, it indicates that
the emergency stop button functions well. Later, restore the emergency stop button and
press the start button, and the screen displays "normal", then the emergency stop button
returns to normal state. If there are abnormalities in the functioning of the emergency
stop button, please contact the technician for consultation and repair.
Storage Conditions and Notice
1)
If the equipment needs to be dismantled before delivery, perform the reverse
installation process.
2)
The system should be stored in a clean and dry place, high-temperature or moist
environment may cause damage to the system parts. If the equipment is out of use
for a long time, it should be stored in the original packaging box.