X-Ray Inspection System
User Manual
P a g e
| 64
Copyright©2021 ZKTECO CO., LTD. All rights reserved.
2)
If it’s not normal, then check the air switch on the equipment to see whether it is
disconnected or not. Check the fuse whether it is disconnected or not. If the fuse is
disconnected, then replace it.
3)
If the connection in the control line socket (DB15) on the special-purpose keyboard
loosens, please re-plug the control line and check the equipment to check whether
it can boot normally.
Failure 2: The conveyor belt is not moving
1)
If it happens, check whether one of the "emergency stop buttons" is pressed
(emergency stop button on the device and console respectively); if so, rotate the
button clockwise and release the button and then press the "Start button" to start
the conveyor belt.
2)
If the conveyor belt still did not run and displays "emergency stop is triggered", then
check the two "micro-switches" on the equipment. Normally, when the two
"sideboards" on the equipment are tightened, the "micro-switches" will be triggered.
If the "micro-switch" is not triggered due to loosening of any one of the "sideboards",
it will also lead to the display message "emergency stop is triggered". Tighten the
sideboard and then press the "Start button" to run the conveyor belt.
3)
If the screen displays "emergency stop is normal", check the "over-current protector"
DZ108-20 of the conveyor whether it trips (means over-current supplied to the
motor cylinder and the trip is caused by over-weight of parcels). If it does, restore it
manually (press the white button on the DZ108-20) and the conveyor must run
normally (For AC motor cylinder).
Failure 3: The screen displays "System self-inspection error"
1)
Situation 1: The system displays "Collector connection error".
This is usually caused by the failure of power supply or transmission lines on the
"transmission board" (inside a small metal box beside the "L-shaped shelf"). Check
the power plug on the "transmission board" to see whether it is loosened or poorly
contacted, and then re-plug the power plug. Check the plug of the transmission line
to see whether it functions well, re-plug this cable plug and the cable on the IPC
board.