ERROR TITLE
ERROR MESSAGE
CAUSE / ACTIONS
Recorder Not
Connected
Confirm the Recorder is within 3
feet of the Tablet, then press the
Pair button on the Recorder. If
there is no flashing green light on
the Recorder, charge the Recorder
before pairing again.
Recorder is out of range of Tablet and/
or Recorder is not charged.
•
•
•
Confirm the Recorder is within 3
feet of the Tablet, then press the Pair
button on the Recorder to re-pair
If there is no flashing green light on
the Recorder, charge the Recorder
before re-pairing
If error continues, contact your
Healthcare Provider
MP Not Connected
Confirm the MP is connected to
the Recorder. The study will resume
automatically after fixing the error.
MP is disconnected from Recorder, not
connected properly or damaged.
•
•
Disconnect/Reconnect MP to Recorder
If error continues, stop the study and
contact your Healthcare Provider
Tablet Unplugged
Confirm Tablet’s power adaptor is
correctly plugged into the Tablet and
an electrical outlet. Study will resume
automatically when error is corrected.
Tablet is not plugged into power,
power adaptor is not connected
properly or electrical outlet is not
functioning.
•
•
•
Confirm power adaptor is correctly
plugged into the Tablet and an
electrical outlet
Confirm the electrical outlet is
functioning
Leave the Tablet plugged in for the
rest of the night to ensure your study
continues
No Air Flow Reading
Confirm the Nasal Cannula is connected
to the Recorder and correctly
positioned in your nose. Blowing your
nose may help. The study will resume
automatically after fixing the error.
Nasal Cannula is not connected to
Recorder, or prongs of Nasal Cannula are
not positioned correctly in your nose, or
your nose is plugged/congested.
•
•
•
•
•
Disconnect/Reconnect Nasal Cannula
to Recorder
Ensure prongs of Nasal Cannula are
properly positioned in each nostril
Check the length of the Nasal Cannula
for kinks
Blow your nose
If error continues, stop the study and
contact your Healthcare Provider
67