4 Installation | 4.3 Installing Additional Components
3. Attach the camera with the adapter to the mi-
croscope's C-mount port.
4. Orient the camera to the stand and fix its position by tightening the clamping screw.
Info
With camera/adapter combinations that have not been recommended expressly by ZEISS, it
may be quite impossible to obtain an unvignetted image.
4.3.9 Replacing Objectives
Procedure
1. Turn the focusing drive to move the mechanical stage down as far as it will go.
2. Turn the nosepiece to move the objective to be changed into a lateral position.
3. Unscrew the objective using the rubber strip in-
cluded in the delivery and remove it down-
ward.
4. Screw the desired objective handtight into the
nosepiece as far as it will go.
5. If you intend to insert the objective in a previously unused lens mount, remove the dust
cover from the corresponding mount of the nosepiece.
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en