4 Installation | 4.3 Installing Additional Components
4.3 Installing Additional Components
4.3.1 Installing or Changing the Tube
Prerequisite
ü
The microscope is unplugged from the mains power.
Procedure
1. Insert the tube to be installed into the stand.
à
The eyepieces must point to the right and its
dovetail mount must be slightly inclined, un-
der the two holding elements.
2. Place the tube horizontally onto the stand.
à
The groove on the underside of the tube must be located above the third holding ele-
ment of the stand.
3. Turn the tube counterclockwise by 90°, align it to the stand.
à
The eyepieces are pointing frontward.
4. Tighten the knurled clamping screw.
5. To detach the tube, loosen the knurled clamping screw.
6. Turn the installed tube clockwise by approximately 90° and detach it on the right side up-
ward.
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en