Plug the ceramic base (8) onto the contact pins of the new halogen
lamp.
•
Insert the new halogen lamp. Make sure you do not touch the lamp
bulb (7) or the interior of the reflector (6).
•
Press the halogen lamp into the spring-loaded mount.
•
Push the lamp module including the new halogen lamp back into the
unit.
•
Press the power switch to turn on the system, and switch on the illu-
mination with the relevant knobs
.
Note:
–
Only use 12 V, 100 W halogen lamps available under Cat. No.:
38 00 79- 9040
To use the service life of your halogen lamps as economically as possible,
we recommend that you proceed as follows: If the main lamp (4) has
failed, remove it and replace it by the backup lamp (5). Install the new hal-
ogen lamp instead of the backup lamp.