Technical data
G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
Electrical data of S81 ceiling mount
Rated voltage
230 V
AC
(220...240 V
AC
± 10%)
115 V
AC
(100...125 V
AC
± 10%)
Current consump-
tion
230 V
AC
3 A
120 V
AC
5 A
100 V
AC
6 A
Rated frequency
50...60 Hz
Fuses
Automatic circuit breaker
Electrical
outlets
–
X-Y coupling
–
Surgical microscope
–
Remote control socket for an external signal
of a maximum of 24 V / 0.5 A.
Electrical
standard
Complying with IEC 60601-1/ EN 60601-1; UL
60601-1;
CAN/CSA-C22.2 No. 601.1-M90
Protection class I, degree of protection IPXO
Type B equipment
Product classification I
as per directive 93/42/EEC, Annex IX
Approval
EMC requirements
The system meets the EMC requirements of
IEC 60601-1-2.
The system meets the RFI requirements of Class
A (hospital grade).
The system has been designed for continuous operation.
1
2
3
1
2
3
Remote socket
View of connector side
176164