G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
Operation
161
Lamp 1: xenon
Lamp 2: halogen
Acoustic signals
Service display
In the event of an error, e.g. during the software check following the
startup of the suspension system, the display and key field displays an
error message in the form of a wrench symbol, accompanied by three
successive beeps.
Motorized control of the surgical microscope is not possible in this case.
With the exception of the recentering of the X-Y coupling, all
other functions of the surgical microscope can only be manually operated.
You can still release the magnetic brakes using the appropriate key in the
left or right handgrip of the surgical microscope.
If you press the "MANUAL" key, the surgical microscope can no longer be
operated via the foot control panel, the handgrips or the display and key
field.
You can continue to use the illumination system. The lamp brightness,
however, is automatically set to a fixed value. You can switch the illumi-
nation on and off on the foot control panel.
Three succes-
sive beeps
–
Error message during the software check after
power-on of the suspension system.
–
Error message in the case of an internal system er-
ror.
One beep
When the focus or zoom position is saved.
One beep
When brightness level 1.0 is reached.
One beep
After power-on of the suspension system.
Intermittent
beep
Error of the illumination system.
MODE
STORE MANUAL