If operated for 8 hours/day on 5 days/week, the lamp needs to be ex-
changed after 12 weeks maximum.
Balance setting
–
Check that the suspension arm has been properly balanced. When
the release button on the surgical microscope is pressed, the effort re-
quired to move the arm up or down must be the same.
Limitation of downward travel
–
The minimum working distance (height) from the surgical field has
been set using the adjustment screw for limiting downward travel.
S88 floor stand only: Stand base
–
Check that the brake tab has been pressed and that the stand is se-
curely locked in position.
Check the accessories
–
Using the manuals provided, check that the other equipment (surgical
microscope, coobservation tube, video system, etc.) is functioning
properly.
Foot control panel
–
Check that the plug of the foot control panel has been connected.
–
Check that the power switch of suspension system has been switched
on.
•
Check that all functions assigned to the respective keys on the foot
control panel are working properly.