134
Preparations
G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
Strain relief device on S88 floor stand
Note:
To prevent inadvertent unplugging of the power cable and of the control
element connector, secure the two cables in strain relief device (1).
After strain relief device (1) has been mounted, the cables must have the
following length:
–
320 mm from the the strain relief device up to and including power out-
let (9).
–
320 mm from the the strain relief device up to connector (10) of the
foot control panel, hand control panel or operating chair equipped with
a footswitch.
•
Form a loop with the cable as shown in (3).
•
Open flap (4).
•
Feed the cable through opening (5).
•
Close flap (6).
•
Tighten the cable until it encloses flap (7).
•
Check the length of the cable.
1