G-30-1529-en
OPMI® VISU 160 on S8, S81 & S88 Suspension Systems
Issue 5.0
Printed on 09. 10. 2006
Maintenance / Further information
189
Motorized focusing and
zoom functions of surgical
microscope are inoperative.
Manual function is acti-
vated. (Yellow LED above
the key lights up).
Switch off manual function.
-
Suspension arm is in hori-
zontal position and cannot
be moved upwards or down-
wards.
Suspension arm still locked. Pull out locking device and
turn though 180°.
-
In connection with three
successive beeps.
–
Error message during
software check after
power-on of the suspen-
sion system.
–
Error message in the
case of internal system
error.
Manual operation is pos-
sible.
Contact service dept.
-
S88 floor stand only: Stand
wobbles.
Floor not level. Stand base
not appropriately posi-
tioned.
Slightly turn stand base. Ar-
ticulated arm should be po-
sitioned at a right angle with
the tilt axis.
Problem
Possible cause
Remedy
See