Installation
37
EN 20_070_0009I v1.5
Switch on
●
To switch on the device, activate the power switch (
6
, Fig. 9) and then
the Standby button (
3
, Fig. 6). The i.Profiler
plus
software will be
started automatically and the
Customer
window will open.
●
After switching on the device, a functional test of measurement
functions must be completed using the test eye device according to the
i.Profiler
plus
Software description
.
Power switch ON (I) / OFF (0) (
6
, Fig. 9) lights up blue when the
device is ready for operation.
CAUTION - HAZARD DUE TO MEASURING ERRORS
Calibration must be reviewed using the test eye device each day before
beginning measurements with customers. A printout of measured values
may be created and filed for documentation purposes. If measured values
on the test eye device are not within the specified tolerances, no
measurements may be carried out on customers! The device must be
decommissioned immediately (see chapter
Decommissioning
, page 42).