Installation
23
EN 20_070_0009I v1.5
Set-up
1
Recessed grips
Fig. 8: Recessed grips on the bottom of the i.Profiler
plus
●
Set up the device so that no light sources (ceiling lights, spotlights, room
lighting, etc.) are pointed directly at the measurement area or its
surroundings (head rest, measuring head).
●
We recommend using a height adjustable device table.
This makes it easier to correctly position the customer in front of the
i.Profiler
plus
to obtain good measurement results.
●
We recommend a minimum table size of 420 x 600 mm for the
standalone i.Profiler
plus
.
●
We recommend not setting the i.Profiler
plus
on a refraction unit with
fixed headrests. Installing it on a refraction unit without fixed headrests
is possible, with an appropriate minimum table size.
●
Set up the device on a horizontal surface.
●
Connect the device to the power network using the power cable.
CAUTION - PROPERTY DAMAGE
To move the device, always use two persons to lift the device by its recessed
grips (
1
, Fig. 8) on the bottom and not by the display! Transport the device on
a movable base with a brake if possible. Avoid shocks during transportation,
since the device contains components that are sensitive to shocks. Always use
the transportation packaging when transporting the device outside of the
company.