HFA3 Instructions for Use
2660021166131 Rev. A 2018-11
7-7
6. Enable MWL and enter these values in the next screen:
A. Enter EmrSCP under AE title.
B. Enter localhost under Host name.
C. Enter 2112 under Port.
D. Select OK.
7. Ensure that the EMR is integrated into the shared/FTP export and worklist folders before
selecting Test connection on the instrument. A check mark will appear if the connection is
successful.
8. Do not change the default values under Specific settings.
Enabled services display a check mark to the left.
Troubleshooting the Connection
If the connection is unsuccessful, take the following actions:
• Check with your network administrator to verify that the instrument is connected to the
network.
• Retry.
• Make sure the appropriate servers are running.
• The storage provider hard drive(s) may be full. Add more hard drives or make more space
available.
• Refer to the DICOM/EMR system manual for a possible resolution of the problem.
• Call Customer Service.