HFA3 Instructions for Use
2660021166131 Rev. A 2018-11
7-3
Export to a Network Shared Folder
1. Select Settings
> Specific Settings to display the Specific Settings screen (see
).
2. Select Report Output.
3. In the End of Test Settings screen, under Report Output Default, select Print, Export, or Export
and Print.
4. Under Report Export File Type select file types.
5. Report Export Location lists the current export location. Select the triple dot
button to
change the export location from the Select folder screen:
A. Choose the left folder icon to navigate to the shared network folder. Select the mapped
folder.
B. Choose the right folder icon to add a folder to the selected export location.
C. Choose Select.
6. You will return to the End of Test Settings screen.
Patient Folders
You may also automatically create patient folders while exporting reports and XML files to a shared
network folder or USB storage device. Turn On Named Patient Folder in the End of Test Settings
screen. An automatically created patient folder name will consist of the patient’s last name, first
name, and date of birth. The patient ID will be added to the beginning of the folder name if the
appropriate setting is enabled.
Connect to a DICOM/EMR Server
This section contains instructions to connect the HFA3 instrument to a DICOM/EMR system
including FORUM. Once connected, the instrument can display a list of patients who are scheduled
for visual field exams from a worklist. When a patient is selected from this list, all demographic data
are entered automatically into the local HFA3 database.
Connection Settings
This section provides a description of the settings required to connect to a DICOM/EMR system. The
majority of these settings do not affect or must be disabled when connecting to a non-DICOM EMR
system (see
“Connect to a Non-DICOM EMR,” on page 7-5
Local Application Entity Settings
Station name: Typically a unique name for the exam lane or instrument. Enter a total of up to 16
characters and spaces.
AE title: Enter a unique AE title for the instrument. Each instrument must have its own unique AE
title which needs to be registered with the DICOM system. Input a total of up to 16 characters (text
is case-sensitive).
Port: Enter the Local Port number that the DICOM/EMR system connects to.
Automatic MWL update: Set to Enabled to automatically update Modality Work Lists.
Frequency of MWL updates: If Automatic MWL update is enabled, the Modality worklist refresh rate
can be changed to the desired time period.
Remote Application Entities
MWL: Enable this service to be able to import a worklist from the DICOM/EMR system.
Storage: Allows storage to the DICOM server.
Query: Enable this service to allow a patient search to FORUM.
Retrieve: Enable this service to automatically retrieve patient information.