HFA3 Instructions for Use
2660021166131 Rev. A 2018-11
4-15
Perform Kinetic Test
The HFA3 Kinetic feature emulates manual standard Goldmann perimetry on Models 840, 850, and
860.
Overview of patient setup and testing on the HFA3:
1. Select Patient.
2. Test Setup.
3. Set up Patient for Testing.
4. Administer the Test (one or both eyes).
5. Review and Save Results.
Select Patient
To select, add, and change patient information follow the instructions in section
or Suprathreshold Test,” on page 4-1.
Test Setup
This section applies to both Kinetic tests. For further details see
“Preset Kinetic Tests,” on page 4-20
Select the Test
For further details see
“Select the Test,” on page 4-3
Select Eye
Next to Perform test on select from the drop-down menu which eye to test: OD (Right), OS (Left), or
Both (default). Next to Start with select from the drop-down menu with which eye, OD or OS, you
want to start.
Note: The Binocular option appears when Manual Kinetic is chosen as the Test Pattern under Test
Parameters. This option will also appear if a custom kinetic pattern is chosen. See
.
Note
See the feature
Test Parameters
For a new patient, go to Test Parameters.
From the Test Type drop-down menu, select Kinetic and choose the
desired Test Pattern.
If Manual Kinetic is chosen you will manually determine stimuli and
test points during the test.
Reports
Additional Information (Optional)
If desired, enter
IOP
(intraocular pressure),
Pupil Diameter
(available only if Auto Pupil is off), and
Visual Acuity
data.
Proceed to Patient Setup
Select Start Test to proceed.