CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Carl Zeiss
Changing the Fuses
DirectFRAP
52
423635-9001-701
05/2014
5.5
Changing the Fuses
Disconnect the power plug in any case before replacing the fuses.
Axio Observer, HBO, XBO, HXP, Tempmodule S1: see separate operating manual.
5.5.1
Signal Distribution Box
The fuse compartment is combined with the
coupler plug and contains two fuses.
•
Switch off the instrument by means of the
power switch.
•
Disconnect the power plug.
•
Remove the fuse carrier (Fig. 5-4/
1
) by pulling it
towards you. You may use a small screwdriver,
if necessary.
•
Remove the fuses (Fig. 5-4/
2
) from the fuse
holder and insert new fuses (2x), type
T1 A/H/250 V
.
•
Push the fuse holder back into the fuse
compartment as far as it will go.
•
Reconnect the power plug.
5.5.2
Laser Bench Tower
•
Switch off the instrument by means of the
power switch.
•
Disconnect the power plug.
•
Pull the fuse holder (Fig. 5-5/
1
) out of the fuse
compartment.
•
Replace relevant fuses (2x
T6.3 A/H/250 V
;
2
).
•
Insert fuse carrier.
•
Reconnect the power plug.
Fig. 5-4
Changing the fuses of the signal
distribution box
Fig. 5-5
Changing the fuses of the Laser
Bench Tower