2.3.4
Basic Laser Safety Kit for Axio Observer.Z1 Stand
The basic laser safety kit for the Axio Observer.Z1 stand is composed of an eyepiece shutter (factory-
installed or to be installed by Service), laser safety board for the stand, distribution box, and labels.
The distribution box for laser safety (Fig. 2-6/
2
) must be connected to the rear of the stand (see Fig. 3-4)
using a 26-pole sub-D connector (Fig. 2-6/
5
).
The four ports of the distribution box (Fig. 2-6/
3
and
4
) are used to connect the following components:
−
Type A
connection: Laser manipulation slider, TIRF slider, remote control switch for laser operation
−
Type B
connection: Laser safety covers, laser safety incubators, Laser Bench Tower or Rack
In addition to that, the distribution box contains a coupler for a door interlock switch (Fig. 2-6/
1
). This
coupler is provided with a factory-installed crossover connector.
For operation, all ports of the distribution box must be occupied.
1
Coupler for additional door interlock switch
2
Distribution box for laser safety
3
Ports for couplers
Type B
4
Ports for couplers
Type A
5
15-pin sub-D connector
Fig. 2-6
Distribution box for laser safety