OPERATION
Carl Zeiss
Illumination and contrast procedures
Axio Examiner
64
M60-2-0003 e 05/2012
•
Re-adjust the knurled screw at the DIC slider (Figure 4-3/
2
), adjusting the diagonal black strip to the
center of the visual field.
•
Remove the auxiliary microscope, and insert the eyepiece again.
(5)
Setting transmitted light DIC
•
Put an objective suitable for DIC into work position using the shift knob (Figure 4-5/
7
).
•
Swing in the analyzer module at the reflector turret (Figure 4-5/
8
).
•
Swing in a suitable DIC prism I, II or III (number as shown at the condenser revolving disk).
The condenser used must not be equipped with DIC prisms which are installed in combination with
polarizing filters.
•
Switch on the SÉNARMONT polarizer (Figure 4-5/
2
)(polarizer and
λ
/4 plate) in the lock-in position.
•
Set the luminous-field diaphragm (Figure 4-5/
8
) and the aperture stop (Figure 4-5/
6
) according to
KÖHLER's rules.
•
To adjust the optimum contrast for SÉNARMONT DIC, turn the polarizer using the thumb wheel
(Figure 4-5/
3
) of the SÉNARMONT polarizer. The
λ
/4 plate located above needs to be switched on at
all times to ensure that contrasting works according to SÉNARMONT. If the DIC slider is in the proper
center position, darkness is achieved in the SÉNARMONT polarizer lock-in position, i.e., the path
difference is 0 nm in this position.
Summary of Contents for Axio Examiner
Page 1: ...Operating Manual Axio Examiner Research Microscope...
Page 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Page 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Page 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Page 96: ......