CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Carl Zeiss
Trouble shooting
Axio Examiner
86
M60-2-0003 e 05/2012
Problem
Cause
Correction
The 12 V 100 W halogen lamp
does not light even though the
on/off switch is switched on.
Power plug not plugged into the
mains outlet.
Plug the power plug into the mains
outlet, checking both equipment
voltage and line voltage.
The 12 V 100 W halogen lamp is not
installed.
Insert he 12 V 100 W halogen lamp,
see page 37.
The 12 V 100 W halogen lamp is
defective.
Replace the 12 V 100 W halogen
lamp, see page 39.
The mandatory 12 V 100 W halogen
lamp is not used.
Use the mandatory 12 V 100 W
halogen lamp.
Fuses are defective.
Replace the fuses, see page 83.
The 12 V 100 W halogen lamp is
flickering. Its luminosity is not
stable.
The 12 V 100 W halogen lamp is at
the end of its mean service life.
Replace the 12 V 100 W halogen
lamp, see page 39.
AC power cord not installed correctly
or broken.
Connect the AC power cord
correctly, or replace it.
The pins of the 12 V 100 W halogen
lamp are not inserted correctly into
the socket.
Insert the pins of the 12 V 100 W
halogen lamp correctly into the
socket,
see page 39.
The motor-operated lower
microscope part does not react.
The microscope is not connected to its
control box.
Connect the microscope,
see page 46.
The control box is either not switched
on or not connected.
Connect or switch on the control
box, see page 46.
The docking station is not connected. Connect the control cable,
see page 46.
The motors are switched off.
Check whether the blue indicator
light is lit (Figure 1-4/
19
); if not,
switch on the motors again,
see page 73.
Summary of Contents for Axio Examiner
Page 1: ...Operating Manual Axio Examiner Research Microscope...
Page 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Page 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Page 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Page 96: ......