EQUIPMENT DESCRIPTION
Carl Zeiss
Technical data
Axio Examiner
24
M60-2-0003 e 05/2012
Power consumption:
Axio Examiner man. (with external power supply unit SNT 12 V DC 100 W) .................................. 220 VA
Axio Examiner mot. (with examiner control) ................................................................................. 260 VA
Transformer mbq52ac-z for HBO 50
Field of application ........................................................................................................................ indoor
Safety class ............................................................................................................................................ I
Degree of protection ....................................................................................................................... IP 20
Line voltage switch-selectable between ................................. 100, 110, 120, 127 VAC and 230, 240 VAC
Power frequency switch-selectable between ........................................................................ 50 and 60 Hz
Maximum power consumption for operation with HBO 50 ........................................................... 350 VA
External power supply unit 12 V DC 100 W, stabilized
Field of application ........................................................................................................................ indoor
Safety class ............................................................................................................................................ I
Degree of protection ....................................................................................................................... IP 20
Line voltage ............................................................................................................ 100 to 240 V ±10 %
No line voltage adjustment is required.
Power frequency ................................................................................................................... 50 to 60 Hz
Power consumption for operation with HAL 100 .......................................................................... 220 VA
HBO 100 W power supply unit
Field of application ........................................................................................................................ indoor
Safety class ............................................................................................................................................ I
Degree of protection ....................................................................................................................... IP 20
Line voltage ................................................................................................................... 100 to 240 VAC
Power frequency ................................................................................................................... 50 to 60 Hz
Power consumption for operation with HBO 103 ......................................................................... 155 VA
Fuses as per IEC 127
Transformer mbq52ac-z for HBO 50 ..................................................................... 100 V, 127 V: 2x T 4 A
........................................................................................................................ .. 220 V - 240 V: 2x T 2.5 A
External power supply unit 12 V DC 100 W ........................................................... 2x T 5.0 A/H, 5x20 mm
HBO 100 W power supply unit ................................................................................. T 2.0 A/H, 5x20 mm
Examiner control .................................................................................................. 2x T 5.0 A/H, 5x20 mm
Light sources
Halogen lamp ...................................................................................................................... 12 V / 50 W
Light source controllability....................................... continuously variable approx. between 3 and 12 V
Halogen lamp .................................................................................................................... 12 V / 100 W
Light source controllability....................................... continuously variable approx. between 3 and 12 V
Mercury vapor short arc lamp ...................................................................................................... HBO 50
Power consumption for HBO 50 ................................................................................................. 50 W
Summary of Contents for Axio Examiner
Page 1: ...Operating Manual Axio Examiner Research Microscope...
Page 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Page 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Page 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Page 96: ......